Categories
Calibration

Calibration of Karl Fischer Titrator (KF)

 

1) Burette Assembly (Dispenser calibration) :-

 

Sr.No. Test

(Vol. ml)

Specification

(± 1 %)

Observation
Wt. of empty Beaker (W1) g Wt. of empty Beaker + sample (W2) g Wt. of sample W3=(W2-W1) g Sample volume =

[W3 / 0.99707] ml

1 0.5 ml 0.49 ml- 0.51ml        
2 1.0 ml 0.99 ml-1.01 ml        
3 5.0 ml 4.95 ml-5.05ml        
4 10.0 ml 9.9 ml- 10.1ml        
5 25.0 ml 24.75 ml-25.25 ml        

 

Density of water is 0.99707 g/ml at 25°C

  

  1. Water content:

 

  1. Karl Fischer Factor

 

 

Sr.No.

Test Specification

(% RSD)

Observation % RSD Remarks
Qty of Water

W (mg)

Burette Reading

BR (ml)

KF Factor

F =W/ BR

1 KF Factor NMT 2.0 %          
     
     

Mean Factor = F =________________

 

  1. Water content of water:

Sr.No. Test Specification Observation Remarks
Qty of Water

W (mg)

Burette Reading

Y ml

% Water Content

(% w/w)

1 Water

content

98 % w/w – 102 % w/w        

 

2. Water content of Sodium Tartrate :

Sr. No. Test Specification Observation Remarks
Qty of Sodium Tartrate

W (mg)

Burette Reading

Y ml

% Water Content

(% w/w)

1 Water

content

15.35% w/w- 15.97% w/w        
       
       

 

Remark: The result Complies/Does not complies as per acceptance criteria.

 

Performed By: ____________     Checked By: ______________

       Date: _____________           Date: ______________

 

“End of Document”

 

Categories
Calibration

Calibration of IR Moisture Analyzer

Calibration of IR Moisture Analyzer

 

  • PROCEDURE
  • Frequency: Calibrate the instrument quarterly.
  •  Operate the instrument as per operation of Moisture Analyzer SOP.
  • Calibrate with the help of standards weight i.e. 1.0gm, 5.0 gm, 10.0 gm and 20.0gm.
  • Acceptance Criteria: ± 0.1%.
  • Set the required temperature with numerical entries in status display and press “enter”.
  • Open the heating module and place the empty sample pan in the sample chamber. Close the module and zero the instrument.
  • Now place the Di-sodium tartrate in the sample pan about 2.0 gm.
  • Set the Temperature at 120°C.
  • Close the heating module and Press “start” key, automatically drying process will take
  • When drying is complete, deep sound will come. Read and record the moisture content of sample in the display and precedes the same process twice.
  • Open the Heating module, carefully remove the sample holder and discard the sample in the dustbin.
  • Acceptance Criteria:60% to 15.70%
  • Calibration Report of IR Moisture Analyser shall be prepared as per SOP.
  • After the successful completion of calibration. Calibration label shall be affix on the instrument as per SOP.
  • If the Calibration parameters are out of limit affix the Out of Calibration label and Inform to QC Head/Designee.
  • Activity shall be recorded in the Instrument usage log book.

 

  • ABBREVIATION(s)
Abbreviation Full Description
QC Quality Control
SOP Standard Operating Procedure.
QA Quality Assurance
% Percentage
gm gram

 

                        

“End of Document”

Categories
Calibration

Calibration of Gas Chromatography (GC)

Calibration of Gas Chromatography (GC)

 

S. No. TEST PARAMETERS LIMITS RESULTS
1 Precision Area ratio RSD NMT 5.0 %  
RT RSD NMT 1.0 %  
2 Linearity Correlation Coefficient ≥ 0.999

Area response of intercept should be less than 5 % of response that is observed at 100 % of the concentration value.

 
3

 

Carry over Acetone NMT 0.1 %  
Tetrahydrofuran  NMT 0.1 %  
4 Precision with Head Space Area ratio RSD NMT 5.0 %  
RT RSD NMT 1.0 %  
5 Linearity with Head Space Correlation Coefficient ≥ 0.999  
6 Carry over Acetone NMT 0.1 %  
Tetrahydrofuran  NMT 0.1 %  
7 Oven Temperature 100°C ± 3°C  

 

150°C ± 3°C  
230°C ± 3°C  
8 Software Verification It should pass  
     

 

1.0: Precision:           

 Batch No. of Acetone   Make  of Acetone  
Batch No. of Tetrahydrofuran    Make  of Tetrahydrofuran  
Column ID  

Standard stock solution:

Acetone Stock Solution:

_____ml Acetone in 100 ml with water(5 %v/v Acetone)

 

Tetrahydrofuran (THF) Stock Solution:

______ml Tetrahydrofuran in 100 ml with water(5 %v/v Tetrahydrofuran)

 

Preparation Standard solution :

Take ——— (10 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution  to ——— (100 ml) with water.

 

S. No RT of

Acetone

RT of Tetrahydrofuran Area of

Acetone

Area of Tetrahydrofuran Area Ratio
1          
2          
3          
4          
5          
6          
Mean        
Stdev.      
RSD      

 

Acceptance Criteria: RSD of Area ratio NMT 5.0 % and RSD of RT NMT 1.0%

 

2.0: Linearity:

Preparation Standard solution:

 a) Take ——— (5 ml) Acetone standard stock solution and ——— (20 ml) Tetrahydrofuran standard stock solution to ——— (200 ml) with water.

 

b) Take ——— (5 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution to ——— (100 ml) with water.

 

c) Take ——— (10 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution to ——— (100 ml) with water.

 

d) Take ——— (15 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution to ——— (100 ml) with water.

 

e) Take ——— (20 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution to ——— (100 ml) with water.

 

S. No. Area of

Acetone

Area of Tetrahydrofuran Area Ratio Average Area Ratio
1      
     
2        
     
3        
     
4        
     
5        
     

 

Linearity =__________________________

 

Intercept shall not be depart from zero.

Acceptance Criteria: Correlation Coefficient ≥ 0.999

 

3.0: Carry over:  

 

Area ratio of blank after six Injections =___________________

Mean Area of Acetone Standard  = ____________________

Area of blank after six Injections

Carryover =  ———————————————————-

Mean Area of Acetone Standard

Area ratio of blank after six Injections =___________________

Mean Area of THF Standard  = ____________________

Area of blank after six Injections

Carryover =  ———————————————————- x 100

Mean Area of THF Standard

 

Acceptance Criteria:   NMT 0.1%

 

4.0: Precision with Head Space:

Standard stock solution:

Acetone Stock Solution:

——ml Acetone in 100 ml with water(5 %v/v Acetone)

Tetrahydrofuran (THF) Stock Solution:

——ml Tetrahydrofuran in 100 ml with water(5 %v/v Tetrahydrofuran)

 

Preparation Standard solution :

Take ——— (10 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution  to ——— (100 ml) with water.

S. No RT of

Acetone

RT of Tetrahydrofuran Area of

Acetone

Area of Tetrahydrofuran Area Ratio
1          
2          
3          
4          
5          
6          
Mean        
Stdev.      
RSD      

 

Acceptance Criteria: RSD of Area ratio NMT 5.0 % and RSD of RT NMT 1.0%

 

5.0: Linearity with Head space:

Preparation Standard solution :

a) Take ——— (5 ml) Acetone standard stock solution and ——— (20 ml) Tetrahydrofuran standard stock solution to ——— (200 ml) with water.

b) Take ——— (5 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution to ——— (100 ml) with water.

c) Take ——— (10 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution to ——— (100 ml) with water.

d) Take ——— (15 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution to ——— (100 ml) with water.

e) Take ——— (20 ml) Acetone standard stock solution and ——— (10 ml) Tetrahydrofuran standard stock solution to ——— (100 ml) with water.

S. No. Area of

Acetone

Area of Tetrahydrofuran Area Ratio Average Area Ratio
1      
     
2        
     
3        
     
4        
     
5        
     

Linearity =__________________________

 Intercept shall not be depart from zero.

 Acceptance Criteria: Correlation Coefficient ≥ 0.999

 

6.0: Carry over with Head Space:     

 

Area ratio of blank after six Injections =___________________

Mean Area of Acetone Standard  = ____________________

Area of blank after six Injections

Carryover =  ———————————————————-

Mean Area of Acetone Standard

Area ratio of blank after six Injections =___________________

Mean Area of THF Standard  = ____________________

Area of blank after six Injections

Carryover =  ———————————————————- x 100

Mean Area of THF Standard

 

Acceptance Criteria:   NMT 0.1%

 

7.0: Oven temperature

Thermometer ID. No. —————————-   

S. No. Set Temperature Observed Temperature Acceptance Criteria (± 30C)
1 100°C   97°C to 103°C
2 150°C   147°C to 153°C
2 230°C   227°C to 233°C

            

8.0: Software Verification

 _____________________________________________________

 

Result: Calibration Complies/Dose not complies

            

Analyzed By Checked By
 

 

 

 
Sign / Date

 

Sign / Date

 

 

  

“End of Document”

 

Categories
Calibration

Calibration of FTIR Spectroscope

Calibration of FTIR Spectroscope

 

1. Control of Resolution Performance:

Parameters Observations Acceptance Criteria
The difference between the percentage transmittance at the transmission maximum at   2870 cm-1 and that at the transmission maximum at 2849.5 cm-1.   Should be greater than 18%
The difference between the percentage transmittance at the transmission maximum at 1589 cm-1 and that at the transmission maximum at 1583 cm-1.   Should be greater than 10%

 

2. Verification of the Wave number scale:

 

Carry out calibration using standard polystyrene film and verify the following Wave numbers (in cm-1).

 

Std. Wavelength Observed Wavelength (cm-1) Acceptance Criteria
3060.0 cm-1   3059.0 – 3061.0 cm-1
2849.5 cm-1   2848.5 – 2850.5 cm-1
1942.9 cm-1   1941.9 – 1943.9 cm-1
1601.2 cm-1   1600.2 – 1602.2 cm-1
1583.0 cm-1   1582.0 – 1584.0 cm-1
1154.5 cm-1   1153.5 – 1155.5 cm-1
1028.3 cm-1   1027.3 – 1029.3 cm-1

 

   3. Reproducibility of the Wave number:

Wave number (cm-1) No.1(cm-1) No.2(cm-1) Error(cm-1) Tolerance(cm-1)
2849.5       ± 5.0
1601.2       ± 1.0
1028.3       ± 1.0

 

The results Complies / Does not complies as per acceptance criteria.

 

Performed By: ________________  Checked By: ______________

Date: ________________                  Date: ______________

 

“End of Document”

 

Categories
Calibration

Calibration of Friability Apparatus

Calibration of Friability Apparatus

 

  • REVOLUTIONS COUNT CHECK:
  • Switch ON the mains.
  • Press the COUNT Key. The display will show the previously set count and the count led will glow.
  • Set required RPM from 25 to 100 for calibration check.
  • Enter RPM from numeric key pad and press ENTER key to register the value.
  • Press start button to start test.
  • Count the number of rotations per minute using a calibrated tachometer.
  • Compare the counted RPM with the set RPM.
  • Limit: Set RPM ± 1RPM.

 

3.5     REVOLUTIONS TIME CHECK:

3.5.1  Switch ON the mains.

3.5.2  Press the TIMER key. The display will show the previously set time and the time led will glow.

3.5.3  Enter 0 hours 4 minutes 00 seconds from numeric key pad.

3.5.4  Press ENTER key to register the values.

3.5.5  Press START key and start the stop watch simultaneously.

3.5.6  Check and compare the time with the set time and electronic display.

  • Repeat steps 3.5.5 and 3.5.6.

LIMITS: 4 minutes ± 10 seconds.

 

REVOLUTIONS COUNT CHECK                          

Sr. No. Set Revolution Observed Revolutions Acceptance criteria
1. 25   Between 24 and 26
2. 50   Between 49 and 51
3. 100   Between 99 and 101

 

REVOLUTIONS TIME CHECK

Sr. No. Set time in Seconds Observed time in Seconds Acceptance criteria

Seconds

1. 240   Between 230 and 249
2. 240  
3. 240  
Average 240  

 

Remarks: Friability Test Apparatus is working satisfactorily/ Not Satisfactorily.

“End of Document”

 

Categories
Calibration

Calibration of Disintegration Apparatus (DT)

Calibration of Disintegration Apparatus (DT)

 

Sr. No. Parameters Set value Acceptance Criteria Observed Values Complies / Does Not Complies
1. Distance travelled by the basket per minute Basket  A (left side) 55 ± 2 mm    
Basket  B (right side)    
2. Temperature of the fluid in the beakers Basket  A (left side) 37ºC ± 2ºC    
Basket  B (right side)    
3.

 

No. of Cycles per min

(Basket  A)

0th to 01st min 29 to 32 Cycles    
0th to 02nd min 58 to 64 Cycles    
0th to 05th min 145 to 160 Cycles    
No. of Cycles per min

(Basket  B)

0th to 01st min 29 to 32 Cycles    
0th to 02nd min 58 to 64 Cycles    
0th to 05th min 145 to 160 Cycles    
4. Integrity of mesh By visual observation check for damage Mesh should not be damaged    
5. Timer Check 10 Minutes 9 Min. 30 Sec.  to  10 Min. 30 sec.    
20 Minutes 19 Min. 30 Sec. to  20 Min. 30 sec.    
30 Minutes 29 Min. 30 Sec. to  30 Min. 30 sec.    

 

Remark: The result Complies/Does not complies as per acceptance criteria.

 

 

Performed By: ________________  Checked By: ______________

Date: ________________                      Date: ______________

 

 

 

“End of Document”

 

Categories
Calibration

Calibration of Drying Oven

Calibration of Drying Oven

 

 

  • PROCEDURE
  •  Frequency: Calibrate the instrument quarterly.
  •  Operate the instrument as per operation of Drying oven SOP
  • Calibrate with the help of calibrated thermometer at a set temperature of 25°C, 50°C, 100°C & 150°C.
  • Acceptance Criteria: ± 1°C of the set temperature
  • Calibration Report of Dry Oven shall be prepared as per
  • After the successful completion of calibration. Calibration label shall be affix on the instrument as per SOP.
  • If the Calibration parameters are out of limit affix the Out of Calibration label as per Format and Inform to QC Head/Designee.

 

  • ABBREVIATION(s)
Abbreviation Full Description
QC Quality Control
SOP Standard Operating Procedure.
QA Quality Assurance
°C Degree Centigrade

 

           

“End of Document”

 

Categories
Calibration

Calibration of Colony Counter

Calibration of Colony Counter

 

  • PROCEDURE
  • PRECAUTIONS
  • Check the cleaning status of colony counter.
  • Operation of Colony Counter:
  • Connect the cord of the instrument to the mains socket and switch ON the main supply.
  • Switch ON the “POWER” button. The internal light shall be flashed ‘ON’ & the digital screen shall display ‘0000’ figure.
  • Reset the counter if required by pressing reset button.
  • Place a blank petridish without agar media on the graduated glass plate and fix the magnifying lens by focusing on the plate at the desired level.
  • Remove the cap of marker.
  • Press the marker gently on the petridish at random locations.
  • Instrument shall beep on every pressing.
  • Press the marker on the petridish counting manually for a count of 25, and check the count registered on the display. The registered count on the display should be 25.
  • Repeat the same procedure for 50, 100, 150, 200 and 250 counts on another petridishes. The display should register a count of 50, 100, 150, 200 and 250 respectively.
  • Record shall be maintained as per annexure.
  • Frequency: Once in a month.

 

  • ABBREVIATION(s)
Abbreviation Full Description
SOP Standard Operating Procedure
IPA Isopropyl Alcohol

 

 

“End of Document”

 

Categories
Calibration

Calibration of Dissolution Apparatus- Physical

Calibration of Dissolution Apparatus- Physical

 

  1. Temperature Calibration:     

 Acceptance Criteria: 37°C ± 0.5°C

Sr. No. JAR No. Set value Observed Values
1. JAR 1 37oC  
2. JAR 2  
3. JAR 3  
4. JAR 4  
5. JAR 5  
6. JAR 6  
7. JAR 7  
8. JAR 8  
9. JAR 9  
10. JAR 10  
11. JAR 11  
12. JAR 12  

 

  1. Paddle and Basket Wobble:                                                        Acceptance Criteria: For Basket NMT 1 mm                                                                   For Paddle NMT 1 mm
Sr. No. Paddle No. Observed Values for Paddle Wobble Basket No. Observed Values for Basket Wobble
1. Paddle 1   Basket 1  
2. Paddle 2   Basket 2  
3. Paddle 3   Basket 3  
4. Paddle 4   Basket 4  
5. Paddle 5   Basket 5  
6. Paddle 6   Basket 6  
7. Paddle 7   Basket 7  
8. Paddle 8   Basket 8  
9. Paddle 9   Basket 9  
10. Paddle 10   Basket 10  
11. Paddle 11   Basket 11  
12. Paddle 12   Basket 12  

 

 

  1. Jar Centering:                                                                 Acceptance Criteria: NMT 2.0 mm

 

Sr. No. JAR No. Observed Values
1. JAR 1  
2. JAR 2  
3. JAR 3  

 

4. JAR 4  
5. JAR 5  
6. JAR 6  
7. JAR 7  
8. JAR 8  
9. JAR 9  
10. JAR 10  
11. JAR 11  
12. JAR 12  

 

  1. Distance between Bottom edge of Basket (D1) and Paddle (D2):

Acceptance Criteria: 25.0 ± 2.0 mm

Sr. No. JAR No. Distance between Bottom edge of Basket (D1) Distance between Bottom edge of Paddle (D2)
1. JAR 1    
2. JAR 2    
3. JAR 3    

 

4. JAR 4    
5. JAR 5    
6. JAR 6    
7. JAR 7    
8. JAR 8    
9. JAR 9    
10. JAR 10    
11. JAR 11    
12. JAR 12    

 

  1. Timer calibration:
Sr. No. Time

(Minutes)

Acceptance Criteria Observation
Displayed Time External Time
1. 5 min

 

5 min ±10 sec    
2. 10 min

 

10 min ±20 sec    
3. 20 min

 

20 min ±40 sec    
4. 30 min

 

30 min ±60 sec    
5. 60 min

 

60 min ±120 sec    

 

  1. RPM Calibration: Acceptance Criteria: RPM± 4%

1)For Paddle

Sr. No. Paddle No. RPM
20

RPM (±1)

50

RPM(±2)

100

RPM(±4)

150

RPM(±6)

200

RPM(±8)

1. Paddle 1          
2. Paddle 2          
3. Paddle 3          
4. Paddle 4          
5. Paddle 5          
6. Paddle 6          
7. Paddle 7          
8. Paddle 8          
9. Paddle 9          
10. Paddle 10          
11. Paddle 11          
12. Paddle 12          

 

2)For Basket

Sr. No. Basket No. RPM
20

RPM (±1)

50

RPM(±2)

100

RPM(±4)

150

RPM(±6)

200

RPM(±8)

1. Basket 1          
2. Basket 2          
3. Basket 3          
4. Basket 4          
5. Basket 5          
6. Basket 6          
7. Basket 7          
8. Basket 8          
9. Basket 9          
10. Basket 10          
11. Basket 11          
12. Basket 12          

 

  1. Mesh Integrity: Acceptance Criteria: Mesh should not be damaged

 

Sr. No. Basket No. Observation
1. Basket 1  
2. Basket 2  
3. Basket 3  
4. Basket 4  
5. Basket 5  
6. Basket 6  
7. Basket 7  
8. Basket 8  
9. Basket 9  
10. Basket 10  
11. Basket 11  
12. Basket 12  

 

Standard Equipment’s used:

Sr. No. Name of standard equipments Sr. No. / ID No. Date of Calibration Valid upto
1. Thermometer      
2. Wobble meter      
3. Centering device      
4. Depth gauge      
5. Tachometer      
6. Stop Watch      

 

 

Remark: The results Complies/Does not comply as per acceptance criteria.

 

 

Performed By: ________________         Checked By: __________

 Date: ________________                           Date: _____________

 

“End of Document”

Categories
Calibration

Calibration of Dissolution Apparatus- Chemical (PVT)

Calibration of Dissolution Apparatus- Chemical (PVT)

 

Dissolution Conditions:     

Apparatus USP Type-1 (Basket) & USP Type-2 (Paddle)
Dissolution Medium 500ml water
Medium Temperature 37°C ± 2°C
Time 30 Minutes
RPM 50 RPM
Wavelength 242nm

 

Reference Standard Details:

 

Sr. No. Name of Reference Standard Batch No./Lot No. Reference No. Validity
1. Prednisone standard      
2. Prednisone tablets 10 mg      

 

Standard Preparation:

Weigh accurately about ________mg of Prednisone RS in ______ ml volumetric flask, add about _____

ml of absolute ethanol, sonicate to dissolve and dilute up to the mark with water. Pipette out _______ ml of above diluted solution in ______ ml volumetric flask and dilute up to mark with water. Take absorbance at 242nm

Observations:

USP Type-1 (Basket)  

 

 

 

 

 

 

 

 

 

 

 

 

USP Type-2 (Paddle)
Sr. No. Absorbance of sample % Drug Dissolved Sr. No Absorbance of sample % Drug Dissolved
1.     1.    
2.     2.    
3.     3.    
4.     4.    
5.     5.    
6.     6.    
7.     7.    
8.     8.    
9.     9.    
10.     10.    
11.     11.    
12     12    
 GM    GM  
%CV     %CV  

Acceptance criteria :   The geometric mean (GM) and percent coefficient of variation (%CV) should be within the limit as per current certificate lot .

 

1) For Paddle

  1. a) GM=____________ b) % CV=___________

 

2) For Basket

  1. a) GM=____________ b) % CV=___________

 

% Dissolution = Spl Abs x Std Wt.x  _5_ x 500  x  Purity of Std x 100

Std Abs        50           50      1                 100          10

 

 

Remark: The results Complies/Does not complies as per acceptance criteria.

 

 

 

Performed By: ________________    Checked By: _____________

 Date: __________                             Date: ___________________

error: Content is protected !!