Calibration of Disintegration Apparatus (DT)

Calibration of Disintegration Apparatus (DT)

 

Sr. No. Parameters Set value Acceptance Criteria Observed Values Complies / Does Not Complies
1. Distance travelled by the basket per minute Basket  A (left side) 55 ± 2 mm    
Basket  B (right side)    
2. Temperature of the fluid in the beakers Basket  A (left side) 37ºC ± 2ºC    
Basket  B (right side)    
3.

 

No. of Cycles per min

(Basket  A)

0th to 01st min 29 to 32 Cycles    
0th to 02nd min 58 to 64 Cycles    
0th to 05th min 145 to 160 Cycles    
No. of Cycles per min

(Basket  B)

0th to 01st min 29 to 32 Cycles    
0th to 02nd min 58 to 64 Cycles    
0th to 05th min 145 to 160 Cycles    
4. Integrity of mesh By visual observation check for damage Mesh should not be damaged    
5. Timer Check 10 Minutes 9 Min. 30 Sec.  to  10 Min. 30 sec.    
20 Minutes 19 Min. 30 Sec. to  20 Min. 30 sec.    
30 Minutes 29 Min. 30 Sec. to  30 Min. 30 sec.    

 

Remark: The result Complies/Does not complies as per acceptance criteria.

 

 

Performed By: ________________  Checked By: ______________

Date: ________________                      Date: ______________

 

 

 

“End of Document”

 