5.1 |
Sampling Containers: |
5.1.1 |
Use clean self sealing polyethylene bags for solid raw materials. |
|
5.1.2 |
Use black polyethylene bags for all light sensitive raw materials. |
|
5.1.3 |
For liquid materials use clean and dry transparent air tight glass bottles. |
|
5.1.4 |
For microbiology samples use sterile polyethylene bags. |
|
5.2 |
Sampling Devices: |
5.2.1 |
For solid raw materials received in big containers/drums/bags use clean and dry stainless steel sampling rods. |
|
5.2.2 |
For solid raw materials received in small containers /bags use clean and dry stainless steel spatulas. |
|
5.2.3 |
For liquid samples use clean and dry liquid sampler. |
|
5.2.4 |
For microbiology samples use clean and dry stainless steel spatulas mopped with 70% IPA. |
|
5.3 |
Safety Precautions: |
5.3.1 |
Wear Personal Protective Equipments such as nose mask, hand gloves and safety goggles at the time of sampling. |
|
5.3.2 |
Do not enter the sampling booth when UV light is “ON”. |
|
5.3.3 |
Sample only one container at a time in a sampling booth. |
|
5.3.4 |
Do not expose hygroscopic materials for longer time in air as it will absorb the moisture. Cover the samples with double polybag. |
|
5.3.5 |
Do not expose light sensitive materials for longer time to light as it will degrade the sample. Cover the samples with black polybag. |
|
5.4 |
Sampling Procedure: |
5.4.1 |
After receiving GRN from Warehouse Department, verify all the documents like GRN, party COA, supplier challan or invoice. |
|
5.4.2 |
If the documents are correlated to each other, the sampling analyst or supervisor shall enter all details in “Raw Material Inward Register” as per Annexure No. 01 and assign the Analytical Reference Number and plan for sampling. |
|
5.4.3 |
Assign Analytical Reference Number (A. R. No.) as “ARYYNNNN”
Where, ‘AR’ stands for ‘Analytical Raw Material’
‘YY’ stands for ‘Last two digits of the current year’.
‘NNNN’ stands for ‘Serial number i.e. 0001, 0002 and so on’. |
|
5.4.4 |
Accompany necessary sampling aids like -poly bags, tie rope or seal, gloves, nose masks, pen marker etc. and accessories like -sampling rod, spatula, spoon, knife, etc as per the nature of sample and pack size. |
|
5.4.5 |
Enter the sampling booth through main entry with proper gowning and accessories with safety hand gloves and nose mask. |
|
5.4.6 |
Operate sampling booth, as per sampling booth SOP. |
|
5.4.7 |
Ensure the cleaning of area and equipment’s are satisfactory and then switch “ON” the RLAF and UV light of pass box and RLAF. |
|
5.4.8 |
Run the UV light for minimum 30 minutes. |
|
5.4.9 |
After 30 minutes switch off the UV light of RLAF and check the differential pressure of RLAF on magnehelic gauze is between 10 mm – 20 mm as well as environment condition, temperature should be NMT 25°C and humidity should be between 45 % RH to 55 % RH. |
|
5.4.10 |
Verify each container/bag/drum with the documents and its label as per manufacturer COA and also check packing integrity at quarantine area. |
|
5.4.11 |
Arrange the container/bag/drum as per lot /batch number and fill required information in the “Raw Material Sampling and Visual Inspection Report” as per Annexure No. 02. |
|
5.4.12 |
If there is any discrepancy with container/bag/drum label and document, then analyst shall report the same to supervisor or quality control head. |
|
5.4.13 |
After successful verification, move the container/bag/drum inside the sampling booth through pass box one at a time. |
|
5.4.14 |
If the material is having microbiological test then use sterile sampling bags, sterile gloves and sampling devices mopped with 70% IPA for sampling of both microbiological and chemical analysis sample to avoid contamination in the material. |
|
5.4.15 |
The person drawing the sample shall ensure that he stands on either side of the container, thereby preventing obstruction of the airflow current to the container being sampled.
HEPA Filter
Either Position of Person |
|
5.4.16 |
All material shall be taken in to sampling booth through material entry door only by SS trolley and ensure that the door is closed after the entry of container. |
|
5.4.17 |
Bring the container under RLAF and open the container. |
|
5.4.18 |
Draw the sample as per sampling plan for respective raw materials. |
|
5.4.19 |
For Active materials: – Perform 100% sampling as per the specifications of the raw materials. |
|
5.4.20 |
For Inactive materials: – Up to 03 containers, bags or drums, perform 100% sampling and for consignments with 04 to 100 or more containers ,bags or drums, perform sampling by √n+1 method where ‘n’ is the number of containers/bags/drums received. |
|
5.4.21 |
For sample quantity refer the respective raw material specifications. |
|
5.4.22 |
For Identification, store sufficient sample quantity (approximately about 0.5 gm) in separate poly bag with proper labelling and numbering as per container/bag/drum for active materials. |
|
5.4.23 |
For Control/Reference samples, store twice the quantity of analytical sample in separate poly bags with proper labelling as per SOP “Storage of Reference /Control Samples”. |
|
5.4.24 |
Ensure the cleaning of sampling booth and accessories are satisfactory and weighing balance is calibrated before sampling. |
|
5.4.25 |
For operation of balance refer to the SOP displayed near the balance. |
|
5.4.26 |
Tare the balance to zero by keeping sampling polybags on it. |
|
5.4.27 |
Open the container/bag/drum with the help of knife or cutter. Open the knot of the polybag inside and insert the sampling rod, hold the sampling rod in cavity close position while` inserting into the material. |
|
5.4.28 |
When rod reach inside of the material up to ¾ portions, move the inner rod so that cavity will open and powder enter into the cavity; again move it to close the cavity. |
|
5.4.29 |
Remove the rod from the material and collect the required sample quantity. Sample quantity shall be at least twice the analysis quantity or as per mentioned in respective specification. |
|
5.4.30 |
For less sample quantity of material use spatula for sampling. |
|
5.4.31 |
Insert spatula into the material preferably middle position of container/bag/drum and take out the sample as per the respective raw material specification. |
|
5.4.32 |
Light Sensitive material should be sampled under subdue light |
|
5.5 |
Example of 100% Sampling: |
5.5.1 |
If the consignment of Paracetamol contains 80 containers, then sample should be drawn from all 80 containers.
Total sample quantity required (Reference sample + Analysis + Residual sample) is 80.0 gm and 0.5 gm from each container for individual identification test.
Sample Quantity to be drawn from each container = 80÷80 = 1.0 gm + 0.5 g separately for individual identification test.
Therefore total sample quantity becomes 80 + (0.5 x 80 = 40) = 120 gm. |
|
|
Example of √n+1 method Sampling: |
5.5.2 |
If the consignment of Magnesium Stearate contains 100 containers, then by applying √n+1 method, where n = 100, the number of containers to be sampled is 11.
Total sample quantity required (Reference sample + Analysis + Residual sample) is 150.0gm.
Sample Quantity to be drawn from each container = 150÷11 = 13.64 gm. |
|
5.5.3 |
Transfer the sample into the sampling polybag tarred before. Weigh the sample polybag on the balance. |
|
5.5.4 |
Put the container/bag/drum number on the sample polybag and deface the manufacturer’s approved label on the container/bag/drum by making cross mark with black marker and affix “UNDER TEST” (yellow colour) label as per Annexure No. 04 on the container with a stamp “SAMPLED” for those container/bag/drum from which the sample is removed along with container number on each “UNDER TEST” label. |
|
5.5.5 |
Reseal the polybag in the container/bag/drum with nylon thread and tightly close the container/bag/drum sampled. |
|
5.5.6 |
Move all the container/bag/drum of the consignment to the under test area. Affix the “UNDER TEST” label on all the containers/bags/drums. |
|
5.5.7 |
Follow the procedure for the sampling of all the container/bag/drums of one consignment/batch. |
|
5.5.8 |
Mix all sample and prepare a composite sample which is sufficient for analysis and reference sample in sampling booth. |
|
|
Note: Prepare composite sample in such a manner so that it represents the quality of whole consignment or batch. |
5.5.9 |
If Microbiology testing is required then divide the sample quantity in three polybags i.e. Microbiology Sample, Control Sample and Chemical Analysis Sample as per respective Raw Material Specification. |
|
5.5.10 |
Visually take note of physical appearance of material while drawing the sample. |
|
5.5.11 |
Keep the analysis sample with “Raw material sample for analysis’ label as per Annexure No. 03 in a designated place in Quality Control Laboratory. |
|
5.5.12 |
Hand Over the Control/Reference samples to Quality Assurance Department for storage. |
|
5.5.13 |
Fill up the log entries and affix “To be cleaned” label on sampling accessories as well as on sampling booth. |
|
5.5.14 |
For cleaning of sampling booth refer to the SOP No. QC/003 “Operation and Cleaning of Sampling Booth”. |
|
5.6 |
Cleaning of Sampling Devices: |
5.6 1 |
Transfer all used sampling devices to washing area from the Sampling Booth after sampling. |
|
5.6 2 |
Laboratory technician must wear the personal protective equipments such as hand gloves, masks for cleaning of sampling devices like spatula and sampling rod. |
|
5.6 3 |
During washing of the sampling devices, if any residue is visually seen then scrub with nylon brush and rinse with potable water and transfer them in 0.5 % soap solution for 5 minutes and wash thoroughly with potable water and finally with purified water. |
|
5.6 4 |
Finally wipe the sampling devices with lint free cloth moistened with 70 % IPA and dry in the oven at 100°C to 105°C for about one hour. Allow to cool, wrap the sampling devices in clean polyethylene bag and affix the “CLEANED” status label with date and signature. |
|
5.6 5 |
Transfer the sampling devices in sampling tool box or in designated place. |
|
5.6 6 |
Record the Usage and Cleaning activity in “Usage and Cleaning Record. |
|
5.7 |
Analysis, Release or Rejection of Raw Materials: |
5.7 1
|
Supervisor/QC Head shall allocate the testing samples to the analysts as per their job responsibility and specialisation. |
|
5.7 2 |
Refer Standard Testing Procedure before initiating the testing. |
|
5.7 3 |
Give requisition for “Analytical Work Report” to the Quality Assurance Department |
|
5.7 4 |
Refer Standard Testing Procedure to ensure that all required reagents, chemicals, accessories, equipment or instruments are standardised or validated and available in the laboratory. |
|
5.7 5 |
Note: – Use amber colour glassware’s for analysis of light sensitive products. |
|
5.7 6 |
In case of non-availability of chemicals, impurities, working standard, instruments or equipments in the laboratory for any specified test then the sample to be send to Government approved outside testing laboratory. |
|
5.7 7 |
If tests to be carried out as per IHS, copy of the protocol to be enclosed or scan copy to be mailed. |
|
5.7 8 |
Fill in the necessary details on the Analytical Work Report according to ‘Raw Material Inward Register’ and start the analysis as per standard testing procedure and specification. |
|
5.7 9 |
In case of active raw materials identification test should be performed on individual container sample. |
|
5.7 10 |
Ensure that all the instruments are calibrated before use. |
|
5.7 11 |
Refer the SOP of instrument before operation and update the instrument operation logs immediately after use. |
|
5.7 12 |
After the analysis, hand over the Analytical work report to Supervisor/Reviewer analyst for data checking. |
|
5.7 13 |
If Supervisor/Reviewer observed any discrepancy with analysis or analytical data, he / she should consult to Head of the Department for investigation. |
|
5.7 14 |
If all the relevant data on Analytical work report complies as per Pharmacopoeia or IHS requirement or both, prepare “Certificate Of Analysis” as per specified format. |
|
5.7 15 |
Full report shall contain AWR, Manufacturer COA, Sampling inspection report, GRN copy and all the instrument prints and COA prepared. |
|
5.7 16 |
Quality control Head shall verify and approve the COA and release the consignment. |
|
5.7 17 |
After approval of COA, analyst should prepare “APPROVED” labels (Green colourwith approved by signature on it and should affix on the “UNDERTEST” label on all containers in such a way that the under test word on the label should be covered but the details on the under test label should be seen. |
|
5.7 18 |
Residual samples: After completion of analysis of Raw Material sample, remaining sample quantity shall be stored as residual sample. Make inward for Residual samples as per “Raw Material Residual Sample Register”. Residual sample shall be store the sample as per specified condition at designated place in Quality Control Laboratory. |
|
|
Note: Residual sample shall be stored after completion of analysis in designated place and Sample should be stored till one year after date of report approval. Destroy the residual sample after one year. |
5.7 19 |
No need to mention potency/Assay and Water/LOD/LOI on approved label in case of inactive materials. |
|
5.7 20 |
Assign Retest Date on the approved label as per the SOP “Retesting of Raw and Packing materials”. |
|
5.7 21 |
Photocopy of the approved COA shall be provided to the Warehouse Department. |
|
5.7 22 |
Destroy Leftover samples/Residual sample as per SOP “Destruction of Laboratory waste”. |
|
5.7 23 |
During analysis if analyst observes the results out of specification or out of limits, he / she should report to the supervisor or Head of the Department. |
|
5.7 24 |
Supervisor shall initiate investigation as per SOP “Out of Specification” by filling OOS form. |
|
5.7 25 |
After completion of investigation, categorise the observation as analyst error, reagent error, instrument error or material defect. |
|
5.7 26 |
If material has major defect that can affect the product quality or out of respective Pharmacopoeial monograph then reject the material. |
|
5.7 27 |
After detailed investigation prepare a COA for rejected material with a remark as “Material does not complies with the prescribed standards’’ and affix the labels on the material as “REJECTED” (Red colour) with rejected by signature and the reason for rejection. |
|
5.7 28 |
Provide Photocopy of Rejected COA to Warehouse Department. Photocopy is not necessary if software is used for material inventory or control. |
|
5.7 29 |
Reference samples of rejected raw materials shall be destroyed as per SOP “Destruction of Laboratory waste”. |
|
5.7 30 |
Rejected Raw Material consignment shall be send back to the manufacturer/supplier for further disposal. |
|
5.7 31 |
If any raw material is not possible to analyse as per Retest Date due to some reasons then material shall be analysed and released prior to taking into manufacturing. |
|
5.7 32 |
Sampling and analysis window: Sampling shall be done within 10 days from the date of GRN Receipt. Testing shall be done within 30 working days from the date of sampling. If any sample exceeds analysis completion window then impact shall be assessed by filing deviation. |
|
5.7 33 |
Re-sampling: Re-sampling shall be done in case of spillage of sample, contamination of samples, and in case of any discrepancies observed during testing. File an incident report and intimate Warehouse department to arrange the material for resampling with required quantity. QC analyst shall perform resampling as procedure mentioned above. Update the resampling quantity in “Raw Material Inward Register”. |
|
5.8 |
Conditional Release of Raw Materials: |
5.8 1 |
In case of urgent release of any batch of Active or Inactive Raw Material is required, then the Quality Control Head shall intimate the Supervisor or responsible analyst to initiate testing on priority. |
|
5.8 2 |
Quality Control Head shall identify the specific tests to be carried out on priority as per the specification of respective Raw Material for Conditional Release. |
|
5.8 3 |
Specific tests such as Description, Identification, Assay, LOD, Water content shall be performed as per Standard Testing Procedure on priority for Conditional Release. |
|
5.8 4 |
Production Department shall raise the request for Conditional Release by filling “Conditional Release Form” if material is required urgently. |
|
5.8 5 |
Hand over Conditional Release form to QA Department for assessment and necessary approval. |
|
5.8 6 |
Raw Materials can be released conditionally by filling Conditional Release form but the product in which the material is used shall be dispatched only after complete testing and approval of COA. |
|
5.8 7 |
Incase if any raw material having Microbiology test, then material can be released conditionally by performing chemical tests but the product in which the material is used shall be dispatched only after complete testing with micro release and approval of COA. |
|
5.8 8 |
Approved labels shall be pasted only after approval of COA and the date of release shall be the date of approval. |
|
|
|
|