Calibration of Colony Counter

Calibration of Colony Counter

 

  • PROCEDURE
  • PRECAUTIONS
  • Check the cleaning status of colony counter.
  • Operation of Colony Counter:
  • Connect the cord of the instrument to the mains socket and switch ON the main supply.
  • Switch ON the “POWER” button. The internal light shall be flashed ‘ON’ & the digital screen shall display ‘0000’ figure.
  • Reset the counter if required by pressing reset button.
  • Place a blank petridish without agar media on the graduated glass plate and fix the magnifying lens by focusing on the plate at the desired level.
  • Remove the cap of marker.
  • Press the marker gently on the petridish at random locations.
  • Instrument shall beep on every pressing.
  • Press the marker on the petridish counting manually for a count of 25, and check the count registered on the display. The registered count on the display should be 25.
  • Repeat the same procedure for 50, 100, 150, 200 and 250 counts on another petridishes. The display should register a count of 50, 100, 150, 200 and 250 respectively.
  • Record shall be maintained as per annexure.
  • Frequency: Once in a month.

 

  • ABBREVIATION(s)
Abbreviation Full Description
SOP Standard Operating Procedure
IPA Isopropyl Alcohol

 

 

“End of Document”

 