Calibration

Calibration of Dissolution Apparatus- Chemical (PVT)

Calibration of Dissolution Apparatus- Chemical (PVT)

 

Dissolution Conditions:     

Apparatus USP Type-1 (Basket) & USP Type-2 (Paddle)
Dissolution Medium 500ml water
Medium Temperature 37°C ± 2°C
Time 30 Minutes
RPM 50 RPM
Wavelength 242nm

 

Reference Standard Details:

 

Sr. No. Name of Reference Standard Batch No./Lot No. Reference No. Validity
1. Prednisone standard      
2. Prednisone tablets 10 mg      

 

Standard Preparation:

Weigh accurately about ________mg of Prednisone RS in ______ ml volumetric flask, add about _____

ml of absolute ethanol, sonicate to dissolve and dilute up to the mark with water. Pipette out _______ ml of above diluted solution in ______ ml volumetric flask and dilute up to mark with water. Take absorbance at 242nm

Observations:

USP Type-1 (Basket)  

 

 

 

 

 

 

 

 

 

 

 

 

USP Type-2 (Paddle)
Sr. No. Absorbance of sample % Drug Dissolved Sr. No Absorbance of sample % Drug Dissolved
1.     1.    
2.     2.    
3.     3.    
4.     4.    
5.     5.    
6.     6.    
7.     7.    
8.     8.    
9.     9.    
10.     10.    
11.     11.    
12     12    
 GM    GM  
%CV     %CV  

Acceptance criteria :   The geometric mean (GM) and percent coefficient of variation (%CV) should be within the limit as per current certificate lot .

 

1) For Paddle

  1. a) GM=____________ b) % CV=___________

 

2) For Basket

  1. a) GM=____________ b) % CV=___________

 

% Dissolution = Spl Abs x Std Wt.x  _5_ x 500  x  Purity of Std x 100

Std Abs        50           50      1                 100          10

 

 

Remark: The results Complies/Does not complies as per acceptance criteria.

 

 

 

Performed By: ________________    Checked By: _____________

 Date: __________                             Date: ___________________