Calibration of FTIR Spectroscope

Calibration of FTIR Spectroscope

 

1. Control of Resolution Performance:

Parameters Observations Acceptance Criteria
The difference between the percentage transmittance at the transmission maximum at   2870 cm-1 and that at the transmission maximum at 2849.5 cm-1.   Should be greater than 18%
The difference between the percentage transmittance at the transmission maximum at 1589 cm-1 and that at the transmission maximum at 1583 cm-1.   Should be greater than 10%

 

2. Verification of the Wave number scale:

 

Carry out calibration using standard polystyrene film and verify the following Wave numbers (in cm-1).

 

Std. Wavelength Observed Wavelength (cm-1) Acceptance Criteria
3060.0 cm-1   3059.0 – 3061.0 cm-1
2849.5 cm-1   2848.5 – 2850.5 cm-1
1942.9 cm-1   1941.9 – 1943.9 cm-1
1601.2 cm-1   1600.2 – 1602.2 cm-1
1583.0 cm-1   1582.0 – 1584.0 cm-1
1154.5 cm-1   1153.5 – 1155.5 cm-1
1028.3 cm-1   1027.3 – 1029.3 cm-1

 

   3. Reproducibility of the Wave number:

Wave number (cm-1) No.1(cm-1) No.2(cm-1) Error(cm-1) Tolerance(cm-1)
2849.5       ± 5.0
1601.2       ± 1.0
1028.3       ± 1.0

 

The results Complies / Does not complies as per acceptance criteria.

 

Performed By: ________________  Checked By: ______________

Date: ________________                  Date: ______________

 

“End of Document”

 