Calibration of IR Moisture Analyzer

Calibration of IR Moisture Analyzer

 

  • PROCEDURE
  • Frequency: Calibrate the instrument quarterly.
  •  Operate the instrument as per operation of Moisture Analyzer SOP.
  • Calibrate with the help of standards weight i.e. 1.0gm, 5.0 gm, 10.0 gm and 20.0gm.
  • Acceptance Criteria: ± 0.1%.
  • Set the required temperature with numerical entries in status display and press “enter”.
  • Open the heating module and place the empty sample pan in the sample chamber. Close the module and zero the instrument.
  • Now place the Di-sodium tartrate in the sample pan about 2.0 gm.
  • Set the Temperature at 120°C.
  • Close the heating module and Press “start” key, automatically drying process will take
  • When drying is complete, deep sound will come. Read and record the moisture content of sample in the display and precedes the same process twice.
  • Open the Heating module, carefully remove the sample holder and discard the sample in the dustbin.
  • Acceptance Criteria:60% to 15.70%
  • Calibration Report of IR Moisture Analyser shall be prepared as per SOP.
  • After the successful completion of calibration. Calibration label shall be affix on the instrument as per SOP.
  • If the Calibration parameters are out of limit affix the Out of Calibration label and Inform to QC Head/Designee.
  • Activity shall be recorded in the Instrument usage log book.

 