Categories
QC SOP

SOP Title: Operation, Cleaning and Calibration of Fourier Transform Infrared Spectrophotometer

SOP Title: Operation, Cleaning and Calibration of Fourier Transform Infrared Spectrophotometer

  

  1. Objective:

To lay down the procedure for Operation, Cleaning and Calibration of Fourier Transform Infrared Spectrophotometer. 

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of Fourier Transform Infrared Spectrophotometer located in Quality Control Laboratory.

 

  1. Responsibility:
  • Quality Control Department: To prepare and review the SOP. To follow the procedures laid down for Operation, Cleaning and Calibration of Fourier Transform Infrared Spectrophotometer as per this SOP.
  • Engineering Department: To carry out preventive maintenance as per schedule and procedure defined.
  • Quality Assurance Department: To review and approve the SOP and Annexures.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department

 

  1. Procedure:
5.1          Safety Precautions/Instructions:
5.1.1 Ensure all the electrical connections are properly earthed. Ensure the power plug of FTIR is fixed in the socket properly.  
5.1.2 Wear personal protective equipments such as hand gloves, nose mask during operation of instrument.  
5.1.3 Ensure that the ON/OFF switch is working properly by switching “ON” FTIR.             
5.2          Precautions:
5.2.1 Do not put off the instrument mains, it should be kept on continuously for 24 hours. In case of a power failure turn the instrument to power ‘OFF’ position.

Accessories required for IR must be stored in a dry box or desiccators.

 
5.2.2 When the power is resorted put the instrument ‘ON’ using switch on the right hand side of the instrument allow an equilibration time of 2 hours before using the instrument.  
5.2.3 Check the temperature and relative humidity. The temperature should be between15 to 30°C and humidity should not be more than 60 %.  
5.2.4 Use IR Grade solvents and Potassium bromide.  
5.2.5 Interference of CO2 can be minimized by avoiding crowding near IR Instrument.  
5.3        Operational  Procedure:
5.3.1 Check the validity of Calibration before starting any operations.  
5.3.2 To switch on the instrument, connect the plug to 230 Volts stabilized power supply.  
5.3.3 Put ‘ON’ the MAINS switch, the “orange” light on the extreme front side of the instrument will glow. This indicates that the dehumidifier is on and working.  
5.3.4 Turn the power ON of FTIR by pressing the button on extreme front right side of the instrument.  It will show green light glowing on left side of button.  
5.3.5 Turn the power on the computer, it will show green light of computer glowing and monitor will go to initialization mode and runs window program and stops at start window.  
5.3.6 Turn on the printer, it will indicate green light.  
5.3.7 Double click on ‘IR SOLUTION’ on the desktop. It will load IR solution on PC and will open the IR SOLUTION window.  
5.3.8 The window comprises of the following items:

Title bar: The title bar displays the name of the software and selected function tabs.

Menu bar: The menu bar displays the menu names like File, Edit, Measurement, AIM, Graph, Admin, Manipulation1, Manipulation 2, Search, Quant, Macro, Print, 3Dim, Window, Environment, Help.

 
5.3.9 Tool bar: Displays commonly used commands as icons. Instead of selecting the command from a menu, click the desired icon for faster and more convenient operation.

Tree view window: The ‘Tree view window’ displays scan parameters, peak tables, data history etc. saved together with data.

Function tabs: The function tabs changes the operation menus such as Measure, View, Manipulation, Search, Quant, AIM, Print form, Macro.

“Status” window: displays user name, the work history, error messages, etc.

Work space: The work space displays spectra and various data.

 
5.3.10 Check that the instrument is on measure mode. Click on ‘measurement’ key then Click on “INITIALISE”.  
5.3.11 Window showing previous background data will get displayed. It will ask whether to remove the data. Click on Yes. The instrument will get initialized.  
5.3.12 After initialization 2 icons in green colour will appear on the right side of the window which is interface and mirror. It shows that the instrument is ready to use.  
5.3.13 Below these parameters like More, Files, Advanced, Data, Instrument is displayed.  
5.3.14 For routine analysis click on ‘Data’ and set the parameters as follows:

Measuring Mode: % T

Apodization: HAPP- Genzel

No of scans: 10

Resolution: 4.0

Range (cm-1):   Minimum = 400 and Maximum = 2000

 
  Blank and Sample disc preparation with KBr Press:
5.3.15 For operation of KBr Press refer SOP ‘Operation, Cleaning and Calibration of KBr Press’.  
5.3.16 Check and ensure that the KBr Press is clean and suitable for use. If not clean it with soft cloth duster.  
5.3.17 Take approx 1 to 3 mg of the sample and 100 to 300 mg of potassium bromide in clean and dried granite mortar and pestle.  
5.3.18 Grind the potassium bromide with the sample in order to mix it thoroughly.  
5.3.19 Assemble the dye by inserting the barrel into the base; insert the shiny stainless steel disc into the barrel with the polished side up.  
5.3.20 Transfer the powder from the mortar into the barrel with the help of clean and dry stainless steel spatula or butter paper in such a way that the surface of the powder in the barrel is levelled and uniformly spread over the barrel. Now insert the plunger slowly and carefully.  
5.3.21 Place the dye assembly in the centre of the KBr press.  Rotate the wheel on top of the KBr press clockwise until it holds the dye assembly tightly.  
5.3.22 Ensure that the dye assembly is tightly placed between the top and bottom of the KBr press.  
5.3.23 Close the valve present at the side of the pressure gauge in the clockwise direction and increase the pressure by moving the piston to about 10 tones and wait for about 2 minutes.  
5.3.24 Release the pressure gently by opening the same valve in the anticlockwise direction.  
5.3.25 Now rotate the wheel in anticlockwise direction and remove the die assembly carefully.  
5.3.26 Remove the Disc by separating the barrel from the base.  Gently tap the plunger on the inverted side so that the steel pellet and KBr disc can be removed from the top. Check that the pellet prepared is transparent and uniform.  
5.3.27 For preparation of blank KBr disc follow the same procedure without the sample.  
5.3.28 The KBr disc/pellet is now ready to be scanned.  
  Scanning of KBr disc.
5.3.29 Transfer the blank disc into the sample holder with the help of butter paper. Put the sample holder into the sample compartment of the instrument.  
5.3.30 On the left hand side of the screen window showing the dialog box as comment and Data file will be displayed. For measuring the background scan enter ‘BKG’ in comment as well as in data file.  
5.3.31
IR Solution
Please prepare sample compartment for Background scan

OK

Click on “BKG”, then the dialog box will appear as:

 
5.3.32 Click on OK. The background measurement will be started. The progress in measurement and the number of scans will be displayed on the status bar at the left hand corner of the screen.  
5.3.33 After the measurement is over the blank spectrum will be directly saved in the ‘VIEW’ window and the ‘VIEW’ window will automatically open on the screen.  
5.3.34 Remove the sample holder containing blank disc from the sample holder. Put the sample disc into the sample holder and insert it back into the sample compartment.  
5.3.35 In the comment window enter the information related to spectra such as sample name, Batch no., A. R. No. and Container no. (For individual identification test) or composite sample.  
5.3.36 In the ‘Data file’ box, type the file name on which the spectrum is to be saved. The IR solution saves all the measured spectra to the hard disc as soon as measurements are finished.  
5.3.37 Click the [….] button at the right end of the data file to select the destination. Set and enter the destination (hard disk and folder) and the file name here. All the files get saved as “smf “files.  
5.3.38 Put the check mark on ‘Auto increment’ option provided on the right side of the data file name to automatically update the file name by putting a number at the end of the file name in every measurement.  
5.3.39 Put the check mark on ‘View background’ to display the measured power spectrum for confirmation when the background measurement is finished.  
5.3.40 Then click on sample. The sample measurement will be started.  
5.3.41 The progress in measurement and the number of scans will be displayed on the status bar at the left hand corner of the screen.  
5.3.42
Stop Scanning X
 

Scanned 5 of 10
Stop Abort Cancel

During measurement, measurement can be paused and restarted by clicking on “STOP” button in the measurement window. Then the dialog box will open as:

 

 
5.3.43 To stop measurement and display the spectrum: Click the [Stop] button.  
5.3.44 To stop measurement and abandon data acquired: Click the [Abort] button.  
5.3.45 To continue measurement: Click the [Cancel] button.  
5.3.46 After the measurement is over the spectrum will be directly saved in the ‘VIEW’ and the ‘VIEW’ window will automatically open on the screen.  
5.3.47 ‘VIEW’ window will show measured spectrum in the upper and lower windows.  
5.3.48 The upper window is called the “Overview” window and the lower window is called “Zooming” window.  
5.3.49 For overlaying the two spectrums one of sample and other of standard, go to “Window” click on “Join visible” and for splitting the two spectra’s again go to “Window” and click on “Split”.  
5.3.50 To display previous spectra from the hard disk, click on ‘File’ on the menu bar and select ‘open’. Then the dialog box will open.  
5.3.51 On the dialog box select a drive and folder in which a spectrum to be open is saved and then select the file name of the spectrum and click ‘OPEN’.  
5.3.52 To open multi spectra at one time, use the [Shift] or [Ctrl] key.  
  For Creation of Print Report Template:
5.3.53 Go to print form option. Click on template .It will show 4 options i.e. Body, Header, Footer, Sizes.  
5.3.54 Click on each option & select / type the parameters, which are required on report, with the help of the window showing the parameters like date, time, file name etc.  
5.3.55 Then save the template. For printing of the report first select the template and then give the print command to the computer.  
5.3.56 For giving the identification of the spectrum annotations can be given. For that go to ‘Graph’ then click on’ Annotations’.  
5.3.57 Type the batch number of sample or standard and then click on “OK”. The name will appear on the spectrum.  
5.3.58 Update the “Instrument/Equipment Usage Log Book” as per SOP “Calibration of Instruments and Equipments”.  
5.4        Cleaning Procedure:
5.4.1 Clean the sample compartment with soft nylon brush before and after use.  
5.5        Calibration Procedure:
  Frequency:  Quarterly or after carrying out any major maintenance work.
5.5.1 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA Department for calibration formats of instruments on the due date of calibration.  
5.5.2 For issuance of calibration report refer SOP “Control of documents, Preparation, Approval, Issuance and Maintenance.  
5.5.3 Perform the calibration of instrument as per Annexure and procedure mentioned below.  
5.5.4 Calibration of FTIR is performed for below parameters:

1. Control of Resolution Performance

2. Verification of the Wave number scale

3. Reproducibility of the wave number

 
5.5.5 First follow the procedure from point 5.3.1 to 5.3.13 and then proceed.  
5.5.6 For calibration click on ‘Data’ and set the parameters as follows:  
5.5.7 Maximize “IR SOLUTION” window. Set the parameters as follows:

Measuring Mode  : % T

Apodization         :  HAPP- Genzel

No. of scans         :  45

Resolution            : 2.0

Range                  : Minimum = 4700, Maximum = 400

 
5.5.8 On the left hand side of the screen window showing the dialog box as comment and Data file will be displayed.  
5.5.9 For measuring the background scan enter ‘BKG’ in comment as well as in data file. Click on “BKG”, then the dialog box will appear as:

IR Solution
Please prepare sample compartment for Background scan

OK

 
5.5.10 Click on OK. The background measurement will be started. The progress in measurement and the number of scans will be displayed on the status bar at the left hand corner of the screen.  
5.5.11 After the measurement is over the blank spectrum will be directly saved in the ‘VIEW’ window and the ‘VIEW’ window will automatically open on the screen.  
5.5.12 After completing background scan, click on the pull down menu of [Measurement] on the menu bar and select [JP Validation].  
5.5.13 Select ‘Instrument Model” and then click OK.  
5.5.14 Now again window [A] will open, click on [Settings], then window [B] will open, click on [Language].  
5.5.15 Select English language and click OK.  
5.5.16 Again window [A] will open, click on [Settings], then window [B] will open, click on [Change-Beam] and then the following window will open.

Select Change-Beam X
Change-Beam

Unusing  
Use   OK
 
5.5.17 Select Change-Beam Unusing and click OK.  
5.5.18 Change temperature, relative humidity.  
5.5.19 After changing the parameters, click OK.  
5.5.20 Following window will open.

     X

Lab control Macro Language

  OK

Remove sample from sample compartment.

 

 
5.5.21 Remove if any sample is present in the sample compartment and click on    OK.  
5.5.22 Wait till the background scan gets complete.  
5.5.23 After the completion of scanning, following window will open.

Lab control Macro Language X
Set polystyrene film into sample compartment.

OK
 
5.5.24 Insert polystyrene film into sample compartment. Click on OK.  
5.5.25 Wait till the sample gets scanned.  
5.5.26 After the completion of scanning, the results will be automatically printed out.  
5.5.27 The following window will be displayed on the screen.

Result X
Pass   (1) Form of a power spectrum

OK

Pass   (2) Resolution, wave number, accuracy and reproducibility test by polystyrene film.

 
5.5.28 Close all windows. Remove the polystyrene film from the sample compartment and keep it in a safe place.  
5.5.29 Enter the details in calibration report as per Annexure. Ensure that the results obtained, comply with the acceptance criteria.  
5.5.30 Update the “Instrument/Equipment Usage Log Book”, “Annual Calibration Schedule” and “Monthly Calibration Schedule” as per SOP “Calibration of Instruments and Equipments”.  
5.5.31 Affix the “Calibration Status Label” on the Instrument as per SOP “Calibration of Instruments and Equipments”.  
5.5.32 If calibration activity is outsourced, the values are to be transcribed in the Annexure, and external party certificate to be attached.  
5.5.33 If any breakdown occurs, follow procedure as per SOP “Calibration of Instruments and Equipments”.  

 

  1. Definitions / Abbreviations:
  • Definitions:
  • Breakdown: Any activity leading to operation of instrument/equipment other than the set parameters or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.
  • Preventive Maintenance: Maintenance activity performed to ensure that the instrument/equipment will function smoothly and to avoid breakdowns. These activities are performed as per predefined frequency.
  • Abbreviations:

 

Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
FTIR Fourier Transform Infra Red
mg milligram
°C Degree Centigrade
% Percentage
BKG Blank
IR Infra Red
CO2 Carbon dioxide
cm-1 Centimetre inverse
Calc Calculation
KBr Potassium Bromide
i.e. That is
A. R. No. Analytical Reference Number

  

 

Calibration Report

 

  • Control of Resolution Performance:
Parameters Observations Acceptance Criteria
The difference between the percentage transmittance at the transmission maximum at   2870 cm-1 and that at the transmission maximum at 2849.5 cm-1.   Should be greater than 18%
The difference between the percentage transmittance at the transmission maximum at 1589 cm-1 and that at the transmission maximum at 1583 cm-1.   Should be greater than 10%

 

  • Verification of the Wave number scale:

 

Carry out calibration using standard polystyrene film and verify the following Wave numbers (in cm-1).

 

Std. Wavelength Observed Wavelength (cm-1) Acceptance Criteria
3060.0 cm-1   3059.0 – 3061.0 cm-1
2849.5 cm-1   2848.5 – 2850.5 cm-1
1942.9 cm-1   1941.9 – 1943.9 cm-1
1601.2 cm-1   1600.2 – 1602.2 cm-1
1583.0 cm-1   1582.0 – 1584.0 cm-1
1154.5 cm-1   1153.5 – 1155.5 cm-1
1028.3 cm-1   1027.3 – 1029.3 cm-1

 

   3) Reproducibility of the Wave number:

 

Wave number (cm-1) No.1(cm-1) No.2(cm-1) Error(cm-1) Tolerance(cm-1)
2849.5       ± 5.0
1601.2       ± 1.0
1028.3       ± 1.0

 

The results Complies / Does not complies as per acceptance criteria.

 

Performed By: ________________         Date: ________________  Checked By: ______________                   Date: ______________                                              

*Note – Ready to use SOP available in “DOWNLOAD” Section.

Categories
QC SOP

SOP Title: Operation, Cleaning and Calibration of KBr Press Apparatus

 SOP Title: Operation, Cleaning and Calibration of KBr Press Apparatus

  1. Objective:

To lay down the procedure for Operation, Cleaning and Calibration of KBr Press Apparatus.

 

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of KBr Press Apparatus located in Quality Control Laboratory.

 

  1. Responsibility:
  • Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation, Cleaning and Calibration of KBr Press Apparatus as per this SOP.
  • Engineering Department: To carry out preventive maintenance as per schedule and procedure defined.
  • Quality Assurance Department: To review and approve the SOP and Annexures.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department

 

  1. Procedure:

 

    5.1 Safety Precautions/Instructions:
5.1.1 Wear personal protective equipments such as safety glasses, hand gloves during operation of equipment.  
    5.2 Operational  Procedure:
5.2.1 Check the validity of calibration before starting any operations.  
5.2.2 Check and ensure that the KBr Press is clean and suitable for use.  
5.2.3 Take approximately 1 to 3 mg of the sample and 100 to 300 mg of potassium bromide (dried for 105oC for 1 hour) in clean and dried granite mortar and pestle.  
5.2.4 Grind the potassium bromide with the sample in order to mix it thoroughly.  
5.2.5 Insert the small shiny stainless steel round disc into the barrel with the polished side up.  
5.2.6 Assemble the dye by inserting the barrel into the base.  
5.2.7 Transfer the powder from the mortar into the barrel with the help of clean and dry stainless steel spatula or butter paper in such a way that the surface of the powder in the barrel is levelled and uniformly spread over the barrel.  
5.2.8 Now insert the plunger slowly and carefully.  
5.2.9 Place the dye assembly in the centre of the KBr press.  
5.2.10 Rotate the wheel on top of the KBr press in clockwise direction until it holds the dye assembly tightly.  
5.2.11 Ensure that the dye assembly is tightly placed between the top and bottom of the KBr press.  
5.2.12 Close the valve present at the side of the pressure gauge in the clockwise direction and increase the pressure by moving the piston to about 10 tons and wait for about 2 minutes.  
5.2.13 Check the pressure on the pressure gauge.  
5.2.14 Release the pressure gently by opening the same valve in the anticlockwise direction.  
5.2.15 Rotate the wheel on top of the KBr press in anti clockwise direction until the dye assembly can be easily removed.  
5.2.16 Remove the sample pellet by separating the barrel from the base.  
5.2.17 Gently tap the plunger on the inverted side so that the small round disc and sample pellet can be removed from the top.  
5.2.18 Check that the sample pellet prepared is transparent and uniform.  
5.2.19 For preparation of blank KBr pellet follow the same procedure without the sample from point no. 5.2.3 to 5.2.18.  
5.2.20 The KBr or sample pellet is now ready to be scanned.  
    5.3 Cleaning Procedure:
5.3.1 Clean the granite mortar and pestle with lint free cloth or tissue paper before and after use.  
5.3.2 Clean the barrel and base of the dye assembly with lint free cloth or tissue paper before and after use.  
5.3.3 Clean the small round disc and plunger first with tissue paper and then apply grease on all sides of the disc and plunger.  
5.3.4 Wrap the disc and plunger in aluminium foil and keep in desiccator.  
5.3.5 Remove the grease of disc and plunger before use with dry tissue paper and then with carbon tetrachloride solvent.  
5.3.6 Clean the KBr press equipment with lint free cloth before and after use.  
5.4 Calibration Procedure:
  Frequency: Once in a Year
5.4.1 Calibration is carried out by external agency as per SOP “Calibration of Instruments and Equipments”.  
5.4.2 Refer Annexure for calibration test to be performed and acceptance criteria.

Sr. No. Tests Set value

in Tons

Acceptance Criteria
1. Pressure Verification 0 ± 2 %
2 ± 2 %
4 ± 2 %
6 ± 2 %
10 ± 2 %
14 ± 2 %
 
5.4.3 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA department for calibration formats of instruments/equipments on the due date of calibration.  
5.4.4 For issuance of calibration report refer SOP “Control of documents, Preparation, Approval, Issuance and Maintenance.  
5.4.5 Update “Annual Calibration Schedule” and “Monthly Calibration Schedule” as per SOP “Calibration of Instruments and Equipments”.  
5.5 Preventive Maintenance:
5.5.1 Perform the preventive maintenance as per Annexure.  
5.5.2 If preventive maintenance activity is outsourced, the observations are to be transcribed in the Annexure and external party certificate to be attached.  
5.5.3 If any breakdown occurs, follow procedure as per SOP “Calibration of Instruments and Equipments”.  
  1. Definitions /Abbreviations:
  • Definitions:
  • Breakdown: Any activity leading to operation of instrument/equipment other than the set parameters or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.
  • Preventive Maintenance: Maintenance activity performed to ensure that the instrument/equipment will function smoothly and to avoid breakdowns. These activities are performed as per predefined frequency.
  • Abbreviations:

 

Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
PM Preventive Maintenance
KBr Potassium Bromide
mg milligram
No. Number
A.R.No. Analytical Reference Number

 

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

 

Categories
QC SOP

SOP Title: Operation, Cleaning and Calibration of Heating Mantle

SOP Title: Operation, Cleaning and Calibration of Heating Mantle

  1. Objective:

To lay down the procedure for Operation, Cleaning and Calibration of Heating Mantle.

 

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of Heating Mantle located in Quality Control Laboratory.

 

  1. Responsibility:
  • Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation, Cleaning and Calibration of Heating Mantle as per this SOP.
  • Engineering Department: To carry out preventive maintenance as per schedule and procedure defined.
  • Quality Assurance Department: To review and approve the SOP and Annexures.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department.

 

  1. Procedure:

 

     5.1 Safety Precautions/Instructions:
5.1.1 Ensure all the electrical connections are properly earthed. Ensure the power plug of the heating mantle is fixed in the socket properly.  
5.1.2 Wear personal protective equipments such as safety glasses, heat resistant hand gloves during operation of equipment.  
5.1.3 Ensure that the hot plate switch is working properly by switching “ON” the heating mantle.             
5.2 Operation  Procedure:
5.2.1 Check the validity of Calibration before starting any operations.  
5.2.2 Ensure that the surface of the coil is clean & dry.  
5.2.3 Switch ON the main switch of heating mantle.  
5.2.4 The red light indicator on the equipment will glow.  
5.2.5 Adjust the required temperature using the knob provided on the front side of the equipment by rotating clockwise direction.  
5.2.6 Keep the sample solution required for heating on the heating mantle.  
5.2.7 Care should be taken that the solution should not overflow on the coil.  
5.2.8 The green light indicator on the equipment will glow after attaining required temperature.  
5.2.9 After use, switch OFF the main switch and remove the sample solution from the heating mantle.  
5.3 Cleaning Procedure:
5.3.1 Switch OFF the main switch.  
5.3.2 Allow to cool and clean the outer surface with lint free cloth.  
5.4 Calibration Procedure:
  Frequency:  Once in a year by external agency or after carrying out any major maintenance work.
5.4.1 Calibration is carried out by external agency as per SOP “Calibration of Instruments and Equipments”.  
5.4.2 Refer Annexure for calibration test to be performed and acceptance criteria.

Sr. No. Tests Set value Acceptance Criteria
1. Temperature Verification 30oC ±2oC
50oC ±2oC
75oC ±2oC
90oC ±2oC
 
5.4.3 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA department for calibration formats of instruments/equipments on the due date of calibration.  
5.4.4 For issuance of calibration report refer SOP “Control of documents, Preparation, Approval, Issuance and Maintenance.  
5.4.5 If any breakdown occurs, follow procedure as per SOP “Calibration of Instruments and Equipments”.  

 

  1. Definitions /Abbreviations:
  • Definitions:
  • Breakdown: Any activity leading to operation of instrument/equipment other than the set parameters or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.

  

  • Abbreviations:
Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
No. Number
A. R. No. Analytical Reference Number

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

 

Categories
QC SOP

SOP Title: Operation, Cleaning and Calibration of Hot Plate

SOP Title: Operation, Cleaning and Calibration of Hot Plate

 

  1. Objective:

To lay down the procedure for Operation, Cleaning and Calibration of Hot Plate.

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of Hot Plate located in Quality Control Laboratory.

  

  1. Responsibility:
  • Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation, Cleaning and Calibration of Hot Plate as per this SOP.
  • Engineering Department: To carry out preventive maintenance as per schedule and procedure defined.
  • Quality Assurance Department: To review and approve the SOP and Annexures.
  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department.

  1. Procedure:

 

    5.1 Safety Precautions/Instructions:
5.1.1 Ensure all the electrical connections are properly earthed. Ensure the power plug of the hot plate is fixed in the socket properly.  
5.1.2 Wear personal protective equipments such as safety glasses, hand gloves during operation of instrument.  
5.1.3 Ensure that the hot plate switch is working properly by switching “ON” the hot plate.             
5.2 Operation  Procedure:
5.2.1  Check the validity of Calibration before starting any operations.  
5.2.2 Introduction to the Keys

Thermostat knob: To increase or decrease temperature of the hot plate.

Heater knob: position  1 for low,

position 2 for medium

position 3 for high

 
5.2.3 Switch ON the main switch of hot plate.  
5.2.4 The red light indicator on the equipment will glow.  
5.2.5 Ensure that the rectangular surface of the hot plate is clean and dry.  
5.2.6 Keep the sample solution required for heating on the hot plate.  
5.2.7 Adjust the required temperature using the knob provided on the front side of the equipment by rotating clockwise direction.  
5.2.8 The green light indicator on the equipment will glow after attaining required temperature.  
5.2.9 Care should be taken that the solutions should not overflow on the hot plate.  
5.2.10 After use, switch OFF the main switch.  
5.3 Cleaning Procedure:
5.3.1 Switch OFF the main switch.  
5.3.2 Allow to cool and clean with lint free cloth.  
5.4 Calibration Procedure:  
  Frequency:  Once in a year by external agency or after carrying out any major maintenance work.
5.4.1 Calibration is carried out by external agency as per SOP number “Calibration of Instruments and Equipments”.  
5.4.2 Refer Annexure for calibration test to be performed and acceptance criteria.

Sr. No. Tests Set value Acceptance Criteria
1. Temperature Verification 30oC ±2oC
50oC ±2oC
75oC ±2oC
90 oC ±2oC
 
5.4.3 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA department for calibration formats of instruments/equipments on the due date of calibration.  
5.4.4 For issuance of calibration report refer SOP “Control of documents, Preparation, Approval, Issuance and Maintenance.  
5.4.5 If any breakdown occurs, follow procedure as per SOP “Calibration of Instruments and Equipments”.  

  

  1. Definitions / Abbreviations:
  • Definitions:
  • Breakdown: Any activity leading to operation of instrument/equipment other than the set parameters or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.
  • Preventive Maintenance: Maintenance activity performed to ensure that the equipment will function smoothly and to avoid breakdowns. These activities are performed as per predefined frequency.

 

  • Abbreviations:
Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
No. Number
A. R. No. Analytical Reference Number

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

Categories
QC SOP

SOP Title: Operation, Cleaning and Calibration of Centrifuge Apparatus

SOP Title: Operation, Cleaning and Calibration of Centrifuge Apparatus

  1. Objective:

To lay down the procedure for Operation, Cleaning and Calibration of Centrifuge Apparatus.

 

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of Centrifuge Apparatus located in Quality Control Laboratory.

 

  1. Responsibility:
  • Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation, Cleaning and Calibration of Centrifuge Apparatus as per this SOP.
  • Engineering Department: To carry out preventive maintenance as per schedule and procedure defined.
  • Quality Assurance Department: To review and approve the SOP and Annexures.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department.

 

  1. Procedure:

 

    5.1 Safety Precautions/Instructions:
5.1.1 Ensure all the electrical connections are properly earthed. Ensure the power plug of the centrifuge apparatus is fixed in the socket properly.  
5.1.2 Ensure that the ON/OFF switch is working properly by Switching  “ON”

the centrifuge apparatus.

 
    5.2 Operation:
5.2.1 Switch ON the main switch. Red light indicator on the equipment will glow.  
5.2.2 Open the lid of centrifuge apparatus.  
5.2.3 Four-centrifuge tube stands are provided inside the equipment.  
5.2.4 Check that the centrifuge tubes are clean and dry before pouring sample solution in it.  
5.2.5 Put the centrifuge tube containing sample solution in the centrifuge apparatus.  
5.2.6 Ensure that the sample solution should not be more than 80% capacity of the centrifuge tube.  
5.2.7 Place the centrifuge tubes in such positions that the opposite sides are balanced.  
5.2.8 When using two centrifuge tubes they should be placed opposite to each other.  
5.2.9 Turn the main knob to the ‘1’ position in clockwise direction.  
5.2.10 Instrument will start at low speed (1000 rpm). Then increase the speed to position 2 (2000 rpm), 3 (4000 rpm), 4 (8000 rpm), 5 (10000 rpm) or as required.  
5.2.11 After completion, slowly decrease the speed on the knob and then finally bring to OFF position.  
5.2.12 Open the lid of centrifuge machine, remove the centrifuge tube and check for satisfactory results.  
5.2.13 Switch OFF the main switch after completion.  
5.3 Cleaning Procedure:
5.3.1 Clean the centrifuge machine externally and internally with clean lint free cloth.  
5.4 Calibration Procedure:
  Frequency:  Once in a year by external agency or after carrying out any major maintenance work.
5.4.1 Calibration is carried out by external agency as per SOP “Calibration of Instruments and Equipments”.  
5.4.2 Refer Annexure for calibration test to be performed and acceptance criteria.

Sr. No. Tests Knob Set value Acceptance Criteria

(± 5%)

1. RPM Verification 1 1000 1000 ± 50 rpm
2 2000 2000 ± 100 rpm
3 4000 4000 ± 200 rpm
4 8000 8000 ± 400 rpm
5 10000 10000 ± 500 rpm
 
5.4.3 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA department for calibration formats of instrument/equipment on the due date of calibration.  
5.4.4 For issuance of calibration report refer SOP “Control of documents, Preparation, Approval, Issuance and Maintenance.  

 

  1. Definitions / Abbreviations:

 

  • Definitions:

 

  • Breakdown: Any activity leading to operation of instrument/equipment other than the set parameters or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.

 

  • Abbreviations:

  

Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
No. Number
A. R. No. Analytical Reference Number

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

Categories
QC SOP

SOP Title: Operation and Calibration of Micrometer Screw Gauge

SOP Title: Operation and Calibration of Micrometer Screw Gauge

  1. Objective:

To lay down the procedure for Operation and Calibration of Micrometer Screw Gauge.

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of Micrometer Screw Gauge located in Quality Control Laboratory.

 

  1. Responsibility:
  • Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation and Calibration of Micrometer Screw Gauge as per this SOP.
  • Quality Assurance Department: To review and approve the SOP and Annexures.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department.

  1. Procedure:

 

    5.1 Safety Precautions/Instructions:
5.1.1 Wear personal protective equipments such as hand gloves during operation of micrometer screw gauge.  
    5.2 Operation:
5.2.1 Range of the instrument is 0 to 25 mm & least count is 0.001mm.  
5.2.2 There are three scale readings on micrometer screw gauge.  
5.2.3 Main scale range is 0 to 25 mm, circular scale range is 0 to 50 and coincide scale range is 0 to 10.  
5.2.4 Loosen the small screw provided by pushing it right side.  
5.2.5 First adjust the circular scale to the 0 reading.  
5.2.6 0 on the main scale should coincide with the 0 of circular scale reading.  
5.2.7 Then carefully insert the sample into the micrometer screw gauge by rotating the circular scale knob to hold firmly.  
5.2.8 First note down the reading on the main scale.  
5.2.9 Then note down the reading on circular scale which is coinciding with the division on the coincide scale.  
5.2.10 Calculate the thickness of the sample using the formula:

Main scale reading + (circular scale x least count)

 
5.2.11 Example:

If main scale reading is 1 mm and circular scale reading coinciding is 44 then by applying formula, thickness of the sample is,

1 + (44 x 0.001)  = 1.044 mm.

 
5.2.12 Remove the sample and bring the instrument in its initial position.  
5.3 Calibration Procedure:
  Frequency:  Once in a year by external agency.
5.3.1 Calibration is carried out by external agency as per SOP “Calibration of Instruments and Equipments”.  
5.3.2 Refer Annexure for calibration test to be performed and acceptance criteria.

Sr. No. Tests Set value Acceptance Criteria (mm)
1. Gauge size 2.5 mm ± 0.004
5.1 mm ± 0.004
7.7 mm ± 0.004
10.3 mm ± 0.004
12.9 mm ± 0.004
15.0 mm ± 0.004
17.6 mm ± 0.004
20.2 mm ± 0.004
22.8 mm ± 0.004
25.0 mm ± 0.004
 
5.3.3 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA department for calibration formats of instruments/equipments on the due date of calibration.  
5.3.4 For issuance of calibration report refer SOP “Control of documents, Preparation, Approval, Issuance and Maintenance.  
5.3.5 Update “Annual Calibration Schedule” and “Monthly Calibration Schedule” as per SOP No “Calibration of Instruments and Equipments”.  
5.3.6 If any breakdown occurs, follow procedure as per SOP “Calibration of Instruments and Equipments”.  

 

 

  1. Definitions / Abbreviations:
  • Definitions:
  • Breakdown: Any activity leading to operation of instrument/equipment other than the set parameters or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.

 

  • Abbreviations:
Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
A. R. No Analytical Reference Number
No. Number
mm millimetre
PM Preventive Maintenance

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

Categories
QC SOP

SOP Title: Operation and Calibration of Vernier Calliper

SOP Title: Operation and Calibration of Vernier Calliper

 

  1. Objective:

To lay down the procedure for Operation and Calibration of Vernier Calliper.

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of Vernier Calliper located in Quality Control Laboratory.

 

  1. Responsibility:
  • Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation and Calibration of Vernier Calliper as per this SOP.
  • Quality Assurance Department: To review and approve the SOP and Annexures.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department.

 

  1. Procedure:

 

  5.1 Safety Precautions/Instructions:
5.1.1 Wear personal protective equipment’s such as hand gloves during operation of vernier calliper.  
   5.2 Operation:
5.2.1 Introduction to keys:

“ON/OFF” (green colour button) → To ON/OFF the instrument.

“in/mm” (white colour button) → display reading will change from millimetre to inch.

“ZERO” (yellow colour button) → To reset the instrument to zero reading.

“INIT” (Black colour button) → To initiate instrument at zero level from any reading.

5.2.2 Measuring range of the instrument is from 0 to 150 mm or 0 to 6 inch.

Resolution – 0.0005 in./0.01 mm.

Instrumental error  – ±0.0015 in./±0.03 mm.

5.2.3 Vernier calliper can be used to take measurement in mm or inch depending upon the requirement.
5.2.4 Loose the top screw slightly so as to down thumb wheel will move smoothly.
5.2.5 Before putting ON the instrument, ensure that both upper and lower grooves of Vernier calliper should touch each other.
5.2.6 First press the ON/OFF button to ON the instrument, display will show “0.0000 in” if the instrument is set for inch OR display will show ”0.00mm” if the instrument is set for mm.
5.2.7 After putting ON instrument, if display is not showing zero reading, then press the “ZERO” button to set zero reading.
5.2.8 For external measurement such as samples, use the lower side grooves of the instrument.
5.2.9 Carefully insert  the  sample  between lower side of two grooves of  vernier  calliper  and  hold  firmly  such  that  sample  should  not  fall  down.
5.2.10 Then note down the display reading.
5.2.11 For internal measurement use the top grooves of the instrument.  
5.2.12 Carefully insert the sample into the top grooves of vernier calliper and hold firmly such that sample should not fall down.
5.2.13 Then note down the display reading.
5.2.14 After testing is completed, remove the sample from instrument; bring the grooves to initial position by moving the thumb wheel and press the “ZERO” button to set zero reading.
5.2.15 Press the ON/OFF button to OFF the instrument.
5.3 Calibration Procedure:
Frequency:  Once in a year by external agency
5.3.1 Calibration is carried out by external agency as per SOP “Calibration of Instruments and Equipments”.
5.3.2 Refer Annexure for calibration test to be performed and acceptance criteria.
5.3.3 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA department for calibration formats of instruments/equipments on the due date of calibration.
5.3.4 For issuance of calibration report refer SOP “Control of documents, Preparation, Approval, Issuance and Maintenance.
5.3.5 Update “Annual Calibration Schedule” and “Monthly Calibration Schedule” as per SOP “Calibration of Instruments and Equipments”.
5.3.6 If any breakdown occurs, follow procedure as per SOP “Calibration of Instruments and Equipments”.

 

  1. Definitions / Abbreviations:
  • Definitions:
  • Breakdown: Any activity leading to operation of instrument/equipment other than the set parameters or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.
  • Abbreviations:
Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
A. R. No Analytical Reference Number

 

No. Number
mm millimetre

  

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

Categories
QC SOP

SOP Title: Cleaning and Sanitation of Chemical Section

SOP Title: Cleaning and Sanitation of Chemical Section

  1. Objective:

To lay down the procedure for Cleaning and Sanitation to be followed in Chemical Section of Quality Control Department.

  1. Scope:

This SOP is applicable for Cleaning and Sanitation of Chemical Section in Quality Control Department.

 

  1. Responsibility:
  • Quality Control Department: To prepare and review the SOP. To follow the procedures laid down for Cleaning and Sanitization of Chemical Section in Quality Control Department as per this SOP.
  • Engineering Department: To supervise the cleaning of Light Fixtures. To support the maintenance of area and provide required utilities.
  • Quality Assurance Department: To ensure compliance of the SOP and to review and approve the SOP and Annexures.
  • Administration Department: To ensure arrangement of the cleaning personnel, cleaning aids and other necessary material/accessories as per requirement.
  1. Accountability:

Head Quality Control Department, Head Administration Department, Head Quality Assurance    Department.

  1. Procedure:

 

  5.1 Precautions:
5.1.1 During wet cleaning cover the electrical panel, switches and other sensitive components with fresh poly bag.  
5.1.2 Do not flush the water / solvent on the electrical switches or tube lights.  
5.1.3           While mopping squeeze the mop properly to remove the water.  
  5.2 Procedure for Cleaning and Sanitisation of Floor:
5.2.1   Collect the dust/debris, unwanted scrap into the waste bin or polybag.  
5.2.2 Use a clean dry lint free mop/cloth and remove dust/debris.  
5.2.3 Use buckets/containers for preparation of disinfectant solution.  
5.2.4  Clean the area with potable water, allow to dry and then mop with disinfectant solution.  
5.2.5 Rotate the disinfectants every week to avoid the development of resistance by microorganisms.  
5.2.6 Mop the area with cleaning agent solution and disinfect the area using disinfectant solution as per the list of disinfectants, concentration and frequency as shown in the below table.  
 

Frequency Name of disinfectants Concentration of disinfectants Dilution
For Area Cleaning and Sanitation:-

Twice daily.

 

 

Dettol 2.5% v/v 125 ml of Dettol in 4.875 litres of potable water.
Savlon 2.5% v/v 125 ml of Savlon  in 4.875 litres of potable water
 Eteol – H 2.5% v/v 125 ml of Eteol – H in 4.875 litres of potable water.

NOTE: Dettol and Savlon are to be changed every week. If either one is not available Eteol –H can be used.

5.2.7          Record the activity in Annexure ‘Area Cleaning and Sanitation Record’.  
5.2.8          Disinfectant solutions should be freshly prepared and should not be stored overnight.  
5.2.9          Balance quantity of disinfectant solutions must be drained in the sink and the bucket must be washed at the end of the day.  
5.2.10      The empty containers of disinfectants should be transferred to scrap yard after rinsing and defacing the label.  
5.2.11      The mopping accessories must be cleaned at the end of the day.  
   5.3 Cleaning of Glasses:
5.3.1 Remove dust from glass surface using a clean and dry duster.  
5.3.2 Clean the glass frame by spraying 1.0 % soap solution and then wipe it with a clean cloth moistened with potable water.  
5.3.3 Clean the glasses with Colin solution/soap solution wipe with glass wiper.  
  Cleaning of Ceilings and Walls:
5.3.4 De-dust the ceiling and walls with the help of dry mop.  
5.3.5 Clean the walls with  clean cloth dipped/moistened with 1 % soap solution using rubber wiper having sufficiently long handle and then with a clean cloth moistened with potable water and finally with disinfectant solution.  
  Cleaning of  Light Fixtures:
5.3.6 Switch OFF the lights prior to cleaning. Then clean the frame /coverings with a dry cloth in the supervision of Engineering Department.  
5.3.7 Wipe the coverings with a moist clean cloth and then wipe it with a dry clean cloth.  
  Cleaning of working tables:
5.3.8 Clean the working tables using a clean duster moistened with potable water.  
  Cleaning of SOP holders/stands, Telephone and Fire Extinguisher:
5.3.9 Remove dust with a dry cotton cloth. Wipe the surface with the help of a clean dry duster.  
   5.4 Cleaning  Frequency:
                      Area            Frequency of Cleaning
 

Floor Sanitising.

Two times daily

i)  Start of shift

ii) End of shift

 Doors and Windows glasses Once in a week.
 Ceiling, walls, Displayed SOP holders/stands,

Telephone and Fire Extinguisher.

Fortnightly.
Working tables  Once daily.
Note: If required in between, clean the area as per requirement.
  Cleaning of Cleaning Aids/Accessories:
5.4.1 After completion of mopping, clean the mop in housekeeping wash area. Wash the Mop thoroughly with potable water.  
5.4.2 Wash the mops and buckets with soap solution and water daily.  
5.4.3 Rinse with clean water, squeeze and remove the water from the Mop.  
5.4.4 Keep the mop for drying in hanging position on hanger provided in wash area.  
5.4.5 Dispose the mop monthly or when it is shredding the threads.  

 

 

  1. Definitions / Abbreviations:
  • Definitions:
  • Disinfectant: The term Disinfectant refers to Savlon or Dettol or Eteol-H solution at 2.5% v/v (25 ml of disinfectant added in potable water to make the volume up to 1000 ml).

% v/v (50 ml of disinfectant is added in potable water to make the      volume up to 1000 ml.

  • Cleaning Solution: 10 ml of Teepol (soap solution) to be added in potable water and make the volume up to 1000 ml (1.0 % v/v).
  • Abbreviations:
Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
No. Number
% Percentage
v/v volume by volume
ml millilitre

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

Categories
QC SOP

SOP Title: Operation, Cleaning and Calibration of Ultrasonic Cleaner

SOP Title: Operation, Cleaning and Calibration of Ultrasonic Cleaner

  1. Objective:

To lay down the procedure for Operation, Cleaning and Calibration of Ultrasonic Cleaner.

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of Ultrasonic Cleaner located in Quality Control Laboratory.

 

  1. Responsibility:
  • Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation, Cleaning and Calibration of Ultrasonic Cleaner as per this SOP.
  • Engineering Department: To carry out preventive maintenance as per schedule and procedure defined.
  • Quality Assurance Department: To review and approve the SOP and Annexures.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department.

  1. Procedure:

 

5.1 Safety Precautions/Instructions:
5.1.1 Before starting the sonicator, check that the sonicator is not empty.  
5.1.2 Fill up the sonicator with purified water at a level of 1 inch from inside tank top.  
5.1.3 Wear personal protective equipments such as safety glasses, hand gloves during operation of equipment.  
5.1.4 Ensure that the ON/OFF switch is working properly by switching “ON” the Ultrasonic Cleaner.             
5.1.5 Do not touch the water surface, when the sonication is going on.  
  Introduction To Keys
5.1.6 ON/OFF→To Start and to Stop the instrument.

SET TIME-   ∧     →will increase the time of the cycle.

∨    →will decrease the time of the cycle.

OFF/ON HEATER →To ON/OFF the heater.

 
5.2 Operation:
5.2.1 Check the validity of Calibration before starting any operations.  
5.2.2 Switch ON the mains. Switch on “POWER” switch of equipment with the indication of green light comes on.  
5.2.3 In order to obtain the maximum efficiency, degas the purified water in the tank for about five minutes.  
5.2.4 Set the required time cycle in minutes by pressing UP or DOWN keys of set time button.  
5.2.5 UP – (∧ ) key will increase the time of the cycle and DOWN-(∨ ) key  will decrease the time of the cycle.  
5.2.6 Insert samples in the flask or glass beaker to be sonicate on mesh stand and press start button.  
5.2.7 When the display shows zero the equipment will shut – OFF automatically.  
5.2.8 Switch ON the heater by pressing ON/OFF key of the heater.  
5.2.9 The temperature is displayed on the screen.  
5.2.10 Remove the sample from the sonicator.  
5.2.11 Switch OFF the unit and disconnect from main switch.  
5.3 Cleaning Procedure:
5.3.1 Cleaning shall be done at the end of the shift.  
5.3.2 Disconnect from main supply from the unit.  
5.3.3 Drain off the water completely from the tap provided at the backside of the sonicator.  
5.3.4 Clean the chamber with wet cloth to remove debris and finally dry the chamber with lint free cloth.  
5.4 Calibration Procedure:
  Frequency:  Once in a year by external agency.  
5.4.1 Calibration is carried out by external agency as per SOP “Calibration of Instruments and Equipments”.  
5.4.2 Refer Annexure for calibration test to be performed and acceptance criteria.

Sr. No. Tests Set value Acceptance Criteria
1. Temperature Verification 25oC  ± 2oC
30oC ± 2oC
50oC ± 2oC
2. Timer Verification 05 min ± 1min
10 min ± 1min
20 min ± 1min
30 min ± 1min
 
5.4.3 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA department for calibration formats of instruments/equipments on the due date of calibration.  
5.4.4 For issuance of calibration report refer “Control of documents, Preparation, Approval, Issuance and Maintenance.  
5.4.5 If any breakdown occurs, follow procedure as per SOP “Calibration of Instruments and Equipments”.  

 

 

  1. Definitions / Abbreviations:
  • Definitions:
  • Breakdown: Any activity leading to operation of equipment other than the set parameters and/ or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.

 

  • Abbreviations:
Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
mm millimetre
A. R. No. Analytical Reference Number

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

Categories
QC SOP

SOP Title: User Account Control System

SOP Title: User Account Control System

  1. Objective:

To lay down the procedure for User Account Control System of instruments.

  1. Scope:

This procedure is applicable for User Account Control System of instruments in the Quality Control Department.

 

  1. Responsibility:
  • Quality Control Department: To prepare and review the SOP. To follow the procedures laid down for User Account Control System of instruments in Quality Control Department as per this SOP.
  • Quality Assurance Department: To review and approve the SOP and Annexure.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department.

  1. Procedure:

 

 5.1 Creation / Modification of User Account
5.1.1          QC Head / Designee shall authorize the system administrator.  
5.1.2          New User Account / Modification in existing user account shall be created / modified on Receipt of requisition from the immediate supervisor of the user as per as per Annexure.  
5.1.3          QC Head / Designee shall approve the request for assigning privileges to the user.  
5.1.4          After carrying out activities as requested in the form, Administrator shall provide the intimation regarding the account information to immediate supervisor of the user and shall maintain user account record as per Annexure.  
5.1.5          Administrator shall configure / reconfigure the privileges to user as per users approved function / role identified in the request.  
5.1.6          List of user account shall be revised periodically whenever required.  
5.1.7          User type and privileges shall be as per Annexure and shall be updated whenever new Instrument/Software installed.  
5.1.8          Account users shall ensure the security and integrity of their password.  
5.1.9          Administrator shall authorize to change the user password in case of either suspect the password is compromised or users forget his password.  
5.1.10      Users shall be responsible for all activity performed with their Login account.  
5.1.11      User shall not allow to perform any activity with others login account.  
5.1.12      IT persons shall have the privileges for data backup and re-store data only.  
 5.2 Training of User :  
5.2.1          Administrator shall train the user on:

·            Usage of application software and its utilities

·            User privileges.

 
5.2.2              Removal of a user account may take place in any of the following conditions:

Ø  In case account user leaves the organization then user account shall be removed permanently.

 
5.2.3     Immediate supervisor of the user shall authorize a request for removal of user account, initiated by the user and approved by QC Head / Designee as per Annexure.  
5.2.4     Administrator shall remove the user account from system and record of user account shall be maintained accordingly in Annexure.  

 

  1. Definitions / Abbreviations :
  • Definitions :
  • Abbreviations:
Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

error: Content is protected !!