SOP Title: Operation and Calibration of Vernier Calliper

SOP Title: Operation and Calibration of Vernier Calliper

 

  1. Objective:

To lay down the procedure for Operation and Calibration of Vernier Calliper.

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of Vernier Calliper located in Quality Control Laboratory.

 

  1. Responsibility:
  • Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation and Calibration of Vernier Calliper as per this SOP.
  • Quality Assurance Department: To review and approve the SOP and Annexures.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department.

 

  1. Procedure:

 

  5.1 Safety Precautions/Instructions:
5.1.1 Wear personal protective equipment’s such as hand gloves during operation of vernier calliper.  
   5.2 Operation:
5.2.1 Introduction to keys:

“ON/OFF” (green colour button) → To ON/OFF the instrument.

“in/mm” (white colour button) → display reading will change from millimetre to inch.

“ZERO” (yellow colour button) → To reset the instrument to zero reading.

“INIT” (Black colour button) → To initiate instrument at zero level from any reading.

5.2.2 Measuring range of the instrument is from 0 to 150 mm or 0 to 6 inch.

Resolution – 0.0005 in./0.01 mm.

Instrumental error  – ±0.0015 in./±0.03 mm.

5.2.3 Vernier calliper can be used to take measurement in mm or inch depending upon the requirement.
5.2.4 Loose the top screw slightly so as to down thumb wheel will move smoothly.
5.2.5 Before putting ON the instrument, ensure that both upper and lower grooves of Vernier calliper should touch each other.
5.2.6 First press the ON/OFF button to ON the instrument, display will show “0.0000 in” if the instrument is set for inch OR display will show ”0.00mm” if the instrument is set for mm.
5.2.7 After putting ON instrument, if display is not showing zero reading, then press the “ZERO” button to set zero reading.
5.2.8 For external measurement such as samples, use the lower side grooves of the instrument.
5.2.9 Carefully insert  the  sample  between lower side of two grooves of  vernier  calliper  and  hold  firmly  such  that  sample  should  not  fall  down.
5.2.10 Then note down the display reading.
5.2.11 For internal measurement use the top grooves of the instrument.  
5.2.12 Carefully insert the sample into the top grooves of vernier calliper and hold firmly such that sample should not fall down.
5.2.13 Then note down the display reading.
5.2.14 After testing is completed, remove the sample from instrument; bring the grooves to initial position by moving the thumb wheel and press the “ZERO” button to set zero reading.
5.2.15 Press the ON/OFF button to OFF the instrument.
5.3 Calibration Procedure:
Frequency:  Once in a year by external agency
5.3.1 Calibration is carried out by external agency as per SOP “Calibration of Instruments and Equipments”.
5.3.2 Refer Annexure for calibration test to be performed and acceptance criteria.
5.3.3 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA department for calibration formats of instruments/equipments on the due date of calibration.
5.3.4 For issuance of calibration report refer SOP “Control of documents, Preparation, Approval, Issuance and Maintenance.
5.3.5 Update “Annual Calibration Schedule” and “Monthly Calibration Schedule” as per SOP “Calibration of Instruments and Equipments”.
5.3.6 If any breakdown occurs, follow procedure as per SOP “Calibration of Instruments and Equipments”.

 

  1. Definitions / Abbreviations:
  • Definitions:
  • Breakdown: Any activity leading to operation of instrument/equipment other than the set parameters or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.
  • Abbreviations:
Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
A. R. No Analytical Reference Number

 

No. Number
mm millimetre

  

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 