SOP Title-Operation of Total Organic Carbon Analyzer (TOC)
- OBJECTIVE
To lay down a procedure for operation of Total Organic Carbon Analyzer.
- SCOPE
This SOP is applicable for operation of Total Organic Carbon Analyzer located at Quality Control Laboratory.
- RESPONSIBILITY
- QC personal shall be responsible for implementation of procedure.
- QC Head – shall be responsible for compliance of this procedure.
- PROCEDURE :
- Operation Procedure :
- Ensure pre start up checks before operating the equipment.
- Instrument is connected to Main power supply
- Switch on the main power switch located on the rear panel of the instrument
- Ensure the software version no. shall be U – 1.00
- Ensure the sample flow rate value is 20 ± 0.5 ml/min is indicated by the four LED’s lights on the front of the sensor.
- If the flow rate is high , only the top LED ‘s will be ON
- If the flow rate is low , only the bottom LED ‘s will be ON
- Operation Procedure :
- If the flow rate is not in the limit adjust the flow rate.
- Press the Menu
- Scroll the configure Menu and press Enter
- Select Yes by pressing ▲ or ▼ to active set flow rate option. This will initiates the set flow rate mode at the sensor. it will stay in this mode until the user changes ‘ Se Flow Rate: Yes’ back ‘ Se Flow Rate: No’
- Select TOC general parameters and press Enter
- Turn ON the pump module and allow water to flow through the module and TOC sensor in to the cylinder and begin the time measurement using a calibrated stop watch. It is important that the water collection and time start should be match.
- Collect a known volume of water with the volumetric container for a specific time, at least 3 minutes
- Record the TOC results of water analysis report .
- Adjustment of flow rate :
- Turn off power to the pump module and TOC sensor.
- Using the pump module cover tool, carefully open the pump module to expose the driver board on the inside of module.
- Turn the potentiometer either clock wise or anti clock wise, depending on the desired
- Rotating the potentiometer either clockwise by 1/4 turn will increase the flow rate by 2ml/min and anticlockwise by 1/4 turn will reduces the flow rate by 2 ml/min.
- Close the cover of the pump module and check the flow rate after each adjustment until the desired the flow rate is 20 ±0.5 ml/min is achieved.
- Cleaning:
- Daily clean the instrument with dry and lint free cloth.
- ABBREVIATION(s)
Abbreviation | Full Description |
SOP | Standard Operating Procedure |
QC | Quality Control |
TOC | Total Organic Carbon |
*Note – Ready to use SOP available in “DOWNLOAD” Section.