SOP Title-Operation of Total Organic Carbon Analyzer (TOC)

SOP Title-Operation of Total Organic Carbon Analyzer (TOC)

 

  • OBJECTIVE

To lay down a procedure  for  operation of  Total Organic Carbon Analyzer.

  • SCOPE

This SOP is applicable for operation of Total Organic Carbon Analyzer located at Quality Control Laboratory.

 

  • RESPONSIBILITY
  • QC personal shall be responsible for implementation of procedure.
  • QC Head – shall be responsible for compliance of this procedure.

 

  • PROCEDURE :
    • Operation Procedure :
      • Ensure pre start up checks before operating the equipment.
      • Instrument is connected to Main power supply
      • Switch on the main power switch located on the rear panel of the instrument
      • Ensure the software version no. shall be U – 1.00
      • Ensure the sample flow rate value is  20 ± 0.5 ml/min is indicated by the four LED’s  lights on the front of the sensor.
      • If the flow rate is high , only the top LED ‘s will be ON
      • If the flow rate is low , only the bottom LED ‘s will be ON

 

  • If the flow rate is not in the limit adjust the flow rate.
  • Press the Menu
  • Scroll the configure  Menu and press Enter
  • Select Yes by pressing ▲ or ▼ to active set flow rate option. This will initiates the set      flow rate mode at the sensor. it will stay in this mode until the user changes ‘ Se Flow  Rate: Yes’ back ‘ Se Flow Rate: No’
  • Select TOC general parameters and press Enter
  • Turn ON the pump module and allow water to flow through the module and TOC sensor in to the cylinder  and begin the time measurement using a calibrated stop watch. It is    important that the water collection and  time start should be match.
  • Collect a known volume of water with the volumetric container for a specific time, at least 3 minutes
  • Record the TOC results of  water analysis report .
  • Adjustment of flow rate :
  • Turn off power to the pump module and TOC sensor.
  • Using the pump module cover tool, carefully open the pump module to expose the driver board on the inside of module.
  • Turn the potentiometer either clock wise or anti clock wise, depending on the desired
  • Rotating the potentiometer either clockwise by  1/4  turn will increase the flow rate by 2ml/min and anticlockwise by  1/4  turn will reduces the flow rate by 2 ml/min.
  • Close the cover of  the pump module and check the flow rate after each adjustment until the desired the flow rate is 20 ±0.5 ml/min is achieved.
  • Cleaning:
    • Daily clean the instrument with dry and lint free cloth.

 

  • ABBREVIATION(s)
Abbreviation  Full Description 
SOP Standard Operating Procedure
QC Quality Control
TOC Total Organic Carbon

 