Categories
Calibration

Calibration of Tap Density Apparatus

Calibration of Tap Density Apparatus

 

Procedure :-

  • Enter 250 tapings by pressing <1000/100> and after <10/1> key and adjust 00250.
  • Press the <START> key to begin a run.
  • Measure the number of Tap per minute using tachometer.
  • Measure the time required to end the set Tap no. with the help of stopwatch.
  • Enter 1250 tapings by pressing key and adjust 01250.
  • Compare the observed Tap with the set tap.
  • Acceptance Criteria:
Set Tapping Acceptance limit
Tapping Time
250 ± 10 tapping 60 ± 2 Seconds
1250 ± 75 tapping 300 ± 10 Seconds

 

SPEED CHECK              

Sr. No. Set Tapings Observed Tapings Acceptance criteria
1. 250  

 

Between 240 and 260
2. 1250  

 

Between 1175 and 1325

 

TAPINGS TIME CHECK

Sr. No. Set Tapings Observed time in Second Acceptance criteria in

Second

1. 250  

 

Between 58 and 62
2. 1250  

 

Between 280 and 310

 

Remark: Instrument is working satisfactorily / not satisfactorily.

 

 

“End of Document”

 