3.3     TEMPERATURE CALIBRATION:

 

  • Operate the instrument as per the SOP.
  • Place a calibrated thermometer inside the constant temperature oven.
  • Set the temperature at 50°C.
  • After temperature is attained, measure the display temperature and standard thermometer temperature and note the readings in the given format.
  • The observed temperature should be within ± 2°C of set temperature.
  • Similarly set the temperature at 100°C and 150°C and repeat the above steps.

 

Temperature Calibration:-

 

Set

Temperature

(°C)

Observed Temperature by thermometer (°C) Display Temperature (°C) Standard Limits
50°C

 

    48°C – 52°C
100°C

 

    98°C – 102°C
150°C

 

    148°C – 152°C

 

Remarks: Constant temperature oven is working satisfactorily/ Not Satisfactorily.

 