Calibration of Drying Oven
3.3 TEMPERATURE CALIBRATION:
- Operate the instrument as per the SOP.
- Place a calibrated thermometer inside the constant temperature oven.
- Set the temperature at 50°C.
- After temperature is attained, measure the display temperature and standard thermometer temperature and note the readings in the given format.
- The observed temperature should be within ± 2°C of set temperature.
- Similarly set the temperature at 100°C and 150°C and repeat the above steps.
Temperature Calibration:-
Set
Temperature (°C) |
Observed Temperature by thermometer (°C) | Display Temperature (°C) | Standard Limits |
50°C
|
48°C – 52°C | ||
100°C
|
98°C – 102°C | ||
150°C
|
148°C – 152°C |
Remarks: Constant temperature oven is working satisfactorily/ Not Satisfactorily.
“End of Document”