Categories
Calibration

Calibration of Melting Point Apparatus

Calibration of Melting Point Apparatus

 

Procedure:-

3.3.3  Fill the capillary tubes with the testing sample VaN/Alin to give a compact column 4mm to 6mm in height. Place this capillary tube in the hole provided in the instrument.

3.3.4  When the temperature reaches within about 10°C below the expected melting point of substance, reduce the rate of heating Adjust the heating rate to about 1°C per minute.

3.3.5  Note down the temperature range at which the material starts to melt and last solid particle of a substance in the capillary tube passes into the liquid phase.

3.3.6  In the same way, repeat the procedure for Sulphanilamide and caffeine. Dry Caffeine at 100°C for three hours, before use.

3.3.7  The standard melting range of the reference substances is given in following table:

 

Sr. No. Reference Substance Standard Melting range
1. Vanillin 81° – 83°C
2. Sulphanilamide 164.5°-166.5°C
3. Caffeine

(dried at 100°C)

234°-237°C

 

  • Tolerance Limit: All the reference material should melt between their standard melting range.

 

Reference Substances detail:

Sr. No Reference Substance Batch Number Make
1 Vanillin    
2 Sulphanilamide    
3 Caffeine    

 

Observations:

Sr. No. Reference Substance Observed Melting Range Tolerance Limit
1 Vanillin   81° to 83°C
2 Sulphanilamide   164.5° to 166.5°C
3 Caffeine   234° to 237°C

 

Remarks      : Melting Point Apparatus is working Satisfactory/ Not Satisfactory

 

“End of Document”

 

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!