Calibration of Analytical Balance-Monthly
Test for Uncertainty:
Standard Weight as per current certified weight (0.5g): ________g Limit = NMT 0.001
No. of Observations | Weight in gm observed | ( xi– x )2 |
1. | ||
2. | ||
3. | ||
4. | ||
5. | ||
6. | ||
7. | ||
8. | ||
9. | ||
10. |
Σ (xi – x ) 2 =________
SD = √ Σ n ( x i – x)2 =__________
i=1 n -1
SD = 3 X SD of 10 Replicate weighing = ________________
Standard weight= ______________________
SD= Standard Deviation
- Checks For Off Centre and Shift Error:
Standard Weight as per current certified weight (0.05g): _________g
Limit = Error NMT 0.5 mg
Position of Weight
♦1 ♦ 2
♦ 5
♦ 3 ♦ 4
Sr. No. | Weight in gm observed | Error (NMT 0.5 mg) |
1 | ||
2 | ||
3 | ||
4 | ||
5 |
- Sensitivity Check: Acceptance Criteria: + 0.1%
Sr. No. | Nominal Weights in gm | Standard Weights in gm (Certified) | Acceptance Criteria
w.r.t std. weights in gm |
Observation in gm |
1. | 100.0000 g | |||
2. | 50.0000 g | |||
3. | 20.0000 g | |||
4. | 10.0000 g | |||
5. | 5.0000 g | |||
6. | 2.0000 g | |||
7. | 1.0000 g | |||
8. | 0.5000 g | |||
9. | 0.2000 g | |||
10. | 0.1000 g | |||
11. | 0.0500 g | |||
12. | 0.0200 g |
The results Complies / Does not complies as per acceptance criteria.
Performed By: ________________ Checked By:__________
Date: ________________ Date: _________