Calibration of HPLC
-
Calibration of Pumps:-
1.1 Flow Accuracy and Flow precision
Observation table for Accuracy of Flow rate
Sr. No. | Set flow rate in
mL/ min |
Time require to collect 10mL volume | Observed flow rate in
mL/ min |
Acceptance criteria
|
|||
1 | 2 | 3 | Average | ||||
1 | 0.5 | 0.49-0.51 | |||||
2 | 1.0 | 0.98-1.02 | |||||
3 | 1.5 | 1.47-1.53 | |||||
4 | 2.0 | 1.96-2.04 | |||||
5 | 5.0 | 4.90-5.10 |
Observation Table for Precision of flow rate
Sr. No. | Set flow rate in
mL/ min |
Observed flow rate in mL/min | %RSD | Acceptance criteria
|
|||
1 | 2 | 3 | Average | ||||
1 | 0.5 | ≤ 2.0% | |||||
2 | 1.0 | ≤ 2.0% | |||||
3 | 1.5 | ≤ 2.0% | |||||
4 | 2.0 | ≤ 2.0% | |||||
5 | 5.0 | ≤ 2.0% |
-
1.2 Gradient Accuracy:-
Observation table for Channel A and B
Eluent Composition | Hx | Calculated Bx (%) | Acceptance Criteria |
20% | 19.4-20.6 | ||
40% | 38.8-41.2 | ||
60% | 58.2-61.8 | ||
80% | 77.6-82.4 | ||
100% | 97.0-103.0 |
Height at 100% composition of B (H100) = __________
Observation table for Channel C and D
Eluent Composition | Hx | Calculated Dx (%) | Acceptance Criteria |
20% | 19.4-20.6 | ||
40% | 38.8-41.2 | ||
60% | 58.2-61.8 | ||
80% | 77.6-82.4 | ||
100% | 97.0-103.0 |
Height at 100% composition of D (H100) = __________
Calculation:
The actually obtained mobile phase composition can be calculated from the Hx values by the following formula.
BX or DX (%) = HX (mm)/H100 (mm)————— x (100)
= _____________________
Where:
X = for the programmed step height (20, 40, 60 and 80).
HX = Height of the respective plateaus in mm of B or D
DX = the actual obtained mobile phase composition.
H100 = Height at 100% composition of B or D.
-
Calibration of Auto Injector:-
-
Vial Test
Sr No | Vial Position/ Location | Observation |
1 | Septum pierced/Not pierced | |
2 | Septum pierced/Not pierced | |
3 | Septum pierced/Not pierced | |
4 | Septum pierced/Not pierced | |
5 | Septum pierced/Not pierced |
-
Volume accuracy of injection:-
Instrument/Equipments/Standards used:
Balance No | |
Balance calibration due date | |
Injection Volume | 20µl |
Number of Injections | 10 |
Density of Water at 200C (d) |
Observation Table:
Initial weight of Vial + Water(W1) in gram |
|
Weight of Vial + Water after 10 Injection(W2) in gram |
|
W = (W1-W2) (in gram) |
|
mL of water injected = W/d in mL |
|
mL of water injected in 10 injection = W/d *1000 µl=A |
|
µl per injection=A/10 |
|
Acceptance Criteria:
The amount of water drawn by the auto injector per injection should be in between 19.0 µl and 21.0µl (i.e., ± 1.0 µl of the set value)
-
Injector Linearity:-
Instrument/Equipments/Standards used:
Balance No | |
Balance calibration due date | |
Standard used | |
Standard weight | |
Standard Lot number | |
HPLC Column details |
Preparation of Standard Solution (20 ppm Caffeine):
- Weighed ______ mg of Caffeine in 100 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
- Further dilute 10mL of this solution to 100mL with Mobile Phase.
Observation Table
Sr. No | Injection Volume | Area |
1 | 5 µl | |
2 | 10 µl | |
3 | 20 µl | |
4 | 50 µl | |
5 | 100 µl | |
Correlation coefficient(r2)
(Acceptance criteria r2 ≥ 0.999) |
-
Injector Precision-
Instrument/Equipments/Standards used:
Balance No | |
Balance calibration due date | |
Standard used | |
Standard weight | |
Standard Lot number | |
HPLC Column details |
Preparation of Standard Solution (20 ppm Caffeine):
- Weighed ______ mg of Caffeine in 100 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
- Further dilute 10mL of this solution to 100mL with Mobile Phase.
Observation Table
Sr. No. | Area |
1 |
|
2 |
|
3 |
|
4 |
|
5 |
|
6 |
|
Average Area |
|
RSD (%) |
|
Acceptance Criteria: RSD NMT 2.0 %
-
Carryover Test :-
Instrument/Equipments/Standards used:
Balance No | |
Balance calibration due date | |
Standard used | |
Standard weight | |
Standard Lot number | |
HPLC Column details |
Preparation of Standard Solution (500 ppm Caffeine):
- Weighed ______ mg of Caffeine in 10 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
Observation Table
Name | Area | % Carryover in Blank |
Blank-1 | NA | |
Sample | NA | |
Blank-2 | ||
Sample | NA | |
Blank-3 | ||
Sample | NA | |
Blank |
Acceptance Criteria: NMT 0.02 %
(Area response of blank-2) – (Area response of blank-1)
Carryover % = ——————————————————— X 100
Average peak area of caffeine obtained from standard solution
Carryover % = ———————————————————
-
Calibration of Detector:-
-
Wave Length Accuracy:-
Instrument/Equipments/Standards used:
Preparation of Standard Solution (20 ppm Caffeine):
- Weighed ______ mg of Caffeine in 100 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
- Further diluted 10mL of this solution to 100mL with Mobile Phase.
Observation Table for wavelength 205 nm
Wavelength nm | Response | Wavelength Maxima/ Maximum response at nm | Acceptance Criteria | |
Area | Height | |||
201 |
|
205nm ± 2 nm | ||
202 | ||||
203 | ||||
204 | ||||
205 | ||||
206 | ||||
207 | ||||
208 | ||||
209 |
Observation Table for wavelength 245 nm
Wavelength nm | Response | Wavelength minima/ Minimum response at nm | Acceptance Criteria | |
Area | Height | |||
241 |
|
245nm ± 2 nm | ||
242 | ||||
243 | ||||
244 | ||||
245 | ||||
246 | ||||
247 | ||||
248 | ||||
249 |
Observation Table for wavelength 273 nm
Wavelength nm | Response | Wavelength Maxima/ Maximum response at nm | Acceptance Criteria | |
Area | Height | |||
269 |
|
273nm ± 2 nm | ||
270 | ||||
271 | ||||
272 | ||||
273 | ||||
274 | ||||
275 | ||||
276 | ||||
277 |
-
Detector Linearity:-
Instrument/Equipments/Standards used:
Preparation of Standard Solution (0.5mg/mL Caffeine):
Weighed ______ mg of Caffeine in 50 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
- 0.25mg/mL standard Solution:
Diluted 1mL of stock solution to 20mL with mobile phase
- 0.5mg/mL standard Solution:
Diluted 2mL of stock solution to 20mL with mobile phase
- 100mg/mL standard Solution:
Diluted 4mL of stock solution to 20mL with mobile phase
- 125mg/mL standard Solution:
Diluted 5mL of stock solution to 20mL with mobile phase
- 250mg/mL standard Solution:
Diluted 10mL of stock solution to 20mL with mobile phase
- 500mg/mL standard Solution:
Use stock solution
Observation Table:
Theoretical Conc.
(mg/ ml) |
Actual Conc.
(mg/ ml) |
Area of injection 1 | Area of injection 2 | Mean Area | Correlation Coefficient |
0.025 | |||||
0.050 | |||||
0.100 | |||||
0.125 | |||||
0.250 | |||||
0.500 |
Acceptance Criteria: Correlation coefficient ≥ 0.999
-
Lamp Energy:-
Acceptance Criteria and results for DAD:
Description | Acceptance criteria | Results | Status Pass/Fail |
Lowest intensity in range 190 nm to 220 nm | Not less than 2000 cts. | ||
Lowest intensity in range 221 nm to 350 nm | Not less than 5000 cts | ||
Lowest intensity in range 351 nm to 500 nm | Not less than 2000 cts | ||
Lowest intensity in range 501 nm to 950 nm | Not less than 4000 cts | ||
Highest intensity in range 190 nm to 350 nm | Not more than 450000 cts | ||
Highest intensity in range 700 nm to 950 nm | Not more than 300000 cts | ||
Highest intensity for D2 alpha line | Not more than 1200000 cts |
Acceptance Criteria and results for VWD:
Description | Acceptance criteria | Results | Status Pass/Fail |
Highest intensity | More than 10000 cts | ||
Lowest intensity | More than 200 cts | ||
Average intensity | More than 5000 cts |
-
Calibration of column Compartment:-
Observation Table for Right side Compartment
Set Temperature (in °C) | Displayed Temperature on instrument
( in° C) |
Observed Temperature on thermometer
( in° C) |
Difference between set temp and observed temp on thermometer ( in° C) |
25.0 | |||
40.0 | |||
60.0 | |||
80.0 |
Observation Table for Left side Compartment
Set Temperature (in °C) | Displayed Temperature on instrument
( in° C) |
Observed Temperature on thermometer
( in° C) |
Difference between set temp and observed temp on thermometer ( in° C) |
25.0 | |||
40.0 | |||
60.0 | |||
80.0 |
Acceptance Criteria:
All the observed thermometer temperature values are within ± 2°C of set value.
-
Calibration of Autosampler Temperature:-
Observation Table
Set Temperature (in °C) | Displayed Temperature on instrument
( in° C) |
Observed Temperature on thermometer
( in° C) |
Difference between set temp and observed temp on thermometer ( in° C) |
25.0 | |||
20.0 | |||
10.0 | |||
4.0 |
Acceptance Criteria:
All the observed thermometer temperature values are within ± 2°C of set value.
Remark: System complies/does not comply with respect to above test.
Performed By: Checked By:
Date: Date: