Categories
QC SOP

SOP Title: Operation, Cleaning and Calibration of Vacuum Oven

SOP Title: Operation, Cleaning and Calibration of Vacuum Oven

  1. Objective:

To lay down the procedure for Operation, Cleaning and Calibration of Vacuum Oven.

  1. Scope:

This SOP is applicable for Operation, Cleaning and Calibration of Vacuum Oven located in Quality Control Laboratory.

 

  1. Responsibility:
  • Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation, Cleaning and Calibration of Vacuum Oven as per this SOP.
  • Engineering Department: To carry out preventive maintenance as per schedule and procedure defined.
  • Quality Assurance Department: To review and approve the SOP and Annexures.

 

  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department

 

  1. Procedure:

 

     5.1 Safety Precautions/Instructions:
5.1.1 Ensure all the electrical connections are properly earthed. Ensure the power plug of the vacuum oven is fixed in the socket properly.  
5.1.2 Wear personal protective equipments such as safety glasses, heat resistant gloves during operation of equipment.  
5.1.3 Ensure that the ON/OFF switch is working properly by switching “ON” the vacuum oven.             
5.2 Operation:
5.2.1 Switch ON the main switch.  
5.2.2 Switch ON the ‘ON/OFF’ knob in ON position.  
5.2.3 The main indicator red light will glow and the temperature panel display will show the temperature in oC.  
5.2.4 Set the required temperature on the temperature panel with ‘TO SET’ key.  
5.2.5 Adjust the required temperature with the ‘COARSE’ and ‘FINE’ keys.  
5.2.6 Wait till the vacuum oven attains the set temperature.  
5.2.7 The Temperature indicator green light will glow when the required temperature is reached.  
5.2.8 Open the door of the vacuum oven by rotating the 3 screw knobs in

anti clockwise direction.

 
5.2.9 Place the sample to be dried inside the vacuum oven on a stainless steel plate provided.  
5.2.10 If the sample is placed in LOD bottle or any other closed glass apparatus, open the lid of the bottle and place in the vacuum oven.  
5.2.11 Immediately close the door and tighten the 3 screw knobs tightly.  
5.2.12 To apply pressure to the vacuum oven, connect the vacuum pump to the vacuum nozzle of the vacuum oven provided at the right side corner of the equipment.  
5.2.13 For operation of vacuum pump refer SOP ‘Operation, Cleaning and Calibration of Vacuum Pump’.  
5.2.14 Open the pressure ‘IN’ knob and start the vacuum pump.  
5.2.15 Check the vacuum gauge dial of the equipment till the needle reaches at the required vacuum.  
5.2.16 As soon as the required vacuum is reached, close the pressure ‘IN’ knob tightly and switch OFF the vacuum pump.  
5.2.17 Allow the substance under test to remain under vacuum at a specified temperature and specified time as mentioned in the respective pharmacopoeia or specification.  
5.2.18 After completion of analysis, slowly release the vacuum by opening the pressure ‘OUT’ knob till the pressure needle on vacuum gauge dial comes to zero position.  
5.2.19 Switch OFF the ‘ON/OFF’ knob in OFF position.  
5.2.20 Open the door of the vacuum oven, remove the sample and place it in dessicator till it attains the room temperature.  
5.2.21 Close the door of the vacuum oven and switch OFF the mains.  
5.2.22 Update the “Instrument/Equipment Usage Log Book” as per SOP “Calibration of Instruments and Equipments”.  
    5.3 Cleaning:  
5.3.1 Switch OFF the mains of the vacuum oven.  
5.3.2 Clean the inside and outside of vacuum oven with lint free cloth before and after use.  
    5.4 Calibration Procedure:
  Frequency:  Once in a year
5.4.1 Calibration is carried out by external agency as per SOP “Calibration of Instruments and Equipments”.  
5.4.2 Refer Annexure for calibration test to be performed and acceptance criteria.

 

Sr. No. Tests Set value Acceptance Criteria

(± 2%)

1. Temperature Verification 50oC ± 1oC
100oC ± 2oC
150oC ± 3oC
200o C ± 4oC
280oC ± 6oC

 

Sr. No. Tests Set value Acceptance Criteria
1. Pressure Verification of vacuum gauge 0 inHg ± 2 %
– 5 inHg  ± 2 %
– 10 inHg ± 2 %
– 15 inHg ± 2 %
– 20 inHg ± 2 %
– 25 inHg ± 2 %
 
  5.4.3 Generate the A. R. No. on the “Instrument/Equipment Calibration Register” as per SOP and give requisition to QA department for calibration formats of instruments/equipments on the due date of calibration.  
   5.4.4 For issuance of calibration report refer SOP “Control of documents, Preparation, Approval, Issuance and Maintenance.  
    5.5 Preventive Maintenance
5.5.1 Perform the preventive maintenance as per Annexure.  
5.5.2 If preventive maintenance activity is outsourced, the observations are to be transcribed in the Annexure, and external party certificate to be attached.  
5.5.3 If any breakdown occurs, follow procedure as per SOP “Calibration of Instruments and Equipments”.  

 

  1. Definitions / Abbreviations :
  • Definitions :
  • Breakdown: Any activity leading to operation of instrument/equipment other than the set parameters or unusual sound or vibration observed in the instrument/equipment.
  • Calibration: Tests performed to ensure that the instrument/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.
  • Preventive Maintenance: Maintenance activity performed to ensure that the equipment will function smoothly and to avoid breakdowns. These activities are performed as per predefined frequency.

 

  • Abbreviations :
Abbreviation Expansion
SOP Standard Operating Procedure
QC Quality Control
QA Quality Assurance
A. R. No Analytical Reference Number
No. Number
oC Degree Celsius
LOD Loss on Drying
PM Preventive Maintenance

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!