Calibration of Viscometer

Calibration of Viscometer

 

 

  • PROCEDURE :-
  • Frequency: Calibrate the instrument quarterly.
  • Operate the instrument as per operation of Viscometer SOP.
  • Level the instrument using leveling screws provided on the base.
  • Turn the power switch located on the rear panel to “ON” position.
  • Calibrate using “500“cps Viscosity Standard fluid:
  • Use Spindle number “S64”.
  • Maintain the temperature of “Viscosity Standard fluid” at 250 C ± 0.10 C, and measure accurately about 16 ml of Viscosity Standard fluid and pour into the sample chamber of UL-ADAPTER.
  • Connect the spindle to lower shaft at the viscometer.
  • Connect the “Sample Chamber” to the mounting channel.
  • Set the RPM to 50 and then press “MOTOR ON/OFF” key. Record the viscosity in cps.
  • Repeat the same operation for two times more.
  • Acceptance Criteria: ± 5% of Certified Value.
  • Calibrate using “100000” cps Viscosity Standard fluid:
  • Use spindle number “S64”.
  • Maintain the temperature of “Viscosity Standard fluid” at 250 C ± 0.10 C, and measure accurately about 16 ml of Viscosity Standard fluid and pour into the sample chamber of UL-ADAPTER.
  • Connect the spindle to lower shaft at the viscometer.
  • Connect the “Sample Chamber” to the mounting channel.
  • Set the RPM to 1.5 and then press “MOTOR ON/OFF” key. Record the viscosity in cps.
  • Repeat the same operation for two times more.
  • Acceptance Criteria: ± 5% of Certified Value.
  • Calibration Report of Viscometer shall be prepared as per Annexure.
  • After the successful completion of calibration. Calibration label shall be affix on the instrument as per SOP.
  • If the Calibration parameters are out of limit affix the Out of Calibration label and Inform to QC Head/Designee.
  • Activity shall be recorded in the Instrument usage log book.

 

  • ABBREVIATION(s)
Abbreviation Full Description
QC Quality Control
SOP Standard Operating Procedure.
QA Quality Assurance
% Percentage
°C Degree Centigrade
RPM Revolution per minute
ml milliliter
cps Centipoise

 

 

           

“End of Document”

 

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!