Calibration of Viscometer
- PROCEDURE :-
- Frequency: Calibrate the instrument quarterly.
- Operate the instrument as per operation of Viscometer SOP.
- Level the instrument using leveling screws provided on the base.
- Turn the power switch located on the rear panel to “ON” position.
- Calibrate using “500“cps Viscosity Standard fluid:
- Use Spindle number “S64”.
- Maintain the temperature of “Viscosity Standard fluid” at 250 C ± 0.10 C, and measure accurately about 16 ml of Viscosity Standard fluid and pour into the sample chamber of UL-ADAPTER.
- Connect the spindle to lower shaft at the viscometer.
- Connect the “Sample Chamber” to the mounting channel.
- Set the RPM to 50 and then press “MOTOR ON/OFF” key. Record the viscosity in cps.
- Repeat the same operation for two times more.
- Acceptance Criteria: ± 5% of Certified Value.
- Calibrate using “100000” cps Viscosity Standard fluid:
- Use spindle number “S64”.
- Maintain the temperature of “Viscosity Standard fluid” at 250 C ± 0.10 C, and measure accurately about 16 ml of Viscosity Standard fluid and pour into the sample chamber of UL-ADAPTER.
- Connect the spindle to lower shaft at the viscometer.
- Connect the “Sample Chamber” to the mounting channel.
- Set the RPM to 1.5 and then press “MOTOR ON/OFF” key. Record the viscosity in cps.
- Repeat the same operation for two times more.
- Acceptance Criteria: ± 5% of Certified Value.
- Calibration Report of Viscometer shall be prepared as per Annexure.
- After the successful completion of calibration. Calibration label shall be affix on the instrument as per SOP.
- If the Calibration parameters are out of limit affix the Out of Calibration label and Inform to QC Head/Designee.
- Activity shall be recorded in the Instrument usage log book.