Calibration of Vacuum Oven
- PROCEDURE:-
- Frequency: Calibrate the instrument quarterly.
- Operate the instrument as per operation of vacuum oven
- Calibrate with the help of calibrated thermometer at a set temperature of 50°, 100° and 150°C.
- Acceptance Criteria: ± 1°C of the set temperature
- Calibration Report of Vacuum Oven shall be prepared as per
- After the successful completion of calibration. Calibration label shall be affix on the instrument as per SOP.
- If the Calibration parameters are out of limit affix the Out of Calibration label and Inform to QC Head/Designee.
- Activity shall be recorded in the Instrument usage log book.
- ABBREVIATION(s)
Abbreviation | Full Description |
QC | Quality Control |
SOP | Standard Operating Procedure. |
QA | Quality Assurance |
°C | Degree Centigrade |
“End of Document”