Calibration of Vacuum Oven

Calibration of Vacuum Oven

 

  • PROCEDURE:-
  • Frequency: Calibrate the instrument quarterly.
  • Operate the instrument as per operation of vacuum oven
  • Calibrate with the help of calibrated thermometer at a set temperature of 50°, 100° and 150°C.
  • Acceptance Criteria: ± 1°C of the set temperature
  • Calibration Report of Vacuum Oven shall be prepared as per
  •  After the successful completion of calibration. Calibration label shall be affix on the instrument as per SOP.
  • If the Calibration parameters are out of limit affix the Out of Calibration label and Inform to QC Head/Designee.
  •  Activity shall be recorded in the Instrument usage log book.

 

  • ABBREVIATION(s)
Abbreviation Full Description
QC Quality Control
SOP Standard Operating Procedure.
QA Quality Assurance
°C Degree Centigrade

 

“End of Document”

 