Calibration

Calibration of UV Cabinet

Calibration of UV Cabinet

 

  • PROCEDURE :-
  •  Frequency: Calibrate the instrument quarterly.
  • Operate the instrument as per operation of UV cabinet SOP.
  • Apply 5µl, 04% w/v solution of sodium salicylate AR and 0.2% w/v solution of sodium Salicylate AR in ethanol (95%), on 10 x 5 cm plate coated with silica gel G and view the spot under wavelength of 254 nm and 365 nm respectively.
  • Acceptance criteria:

Intensity of the spot should be clear and visible of 0.04% w/v solution of sodium salicylate AR and 0.2% w/v solution of sodium Salicylate AR in ethanol (95%) at 254 nm and 365 nm respectively.

  • Calibration Report of UV Cabinet shall be prepared as per Annexure.
  • After the successful completion of calibration. Calibration label shall be affix on the instrument as per SOP.
  • If the Calibration parameters are out of limit affix the Out of Calibration label and Inform   to QC Head/Designee
  • Activity shall be recorded in the Instrument usage log book.

 

  • ABBREVIATION(s)
Abbreviation Full Description
%w/v Percent Wight/Volume
cm Centimeter
UV Ultra-Violet

“End of Document”

 