Calibration of HPLC
-
Calibration of Pumps:-
1.1 Flow Accuracy and Flow precision
Observation table for Accuracy of Flow rate
Sr. No. |
Set flow rate in
mL/ min |
Time require to collect 10mL volume |
Observed flow rate in
mL/ min |
Acceptance criteria
|
1 |
2 |
3 |
Average |
1 |
0.5 |
|
|
|
|
|
0.49-0.51 |
2 |
1.0 |
|
|
|
|
|
0.98-1.02 |
3 |
1.5 |
|
|
|
|
|
1.47-1.53 |
4 |
2.0 |
|
|
|
|
|
1.96-2.04 |
5 |
5.0 |
|
|
|
|
|
4.90-5.10 |
Observation Table for Precision of flow rate
Sr. No. |
Set flow rate in
mL/ min |
Observed flow rate in mL/min |
%RSD |
Acceptance criteria
|
1 |
2 |
3 |
Average |
1 |
0.5 |
|
|
|
|
|
≤ 2.0% |
2 |
1.0 |
|
|
|
|
|
≤ 2.0% |
3 |
1.5 |
|
|
|
|
|
≤ 2.0% |
4 |
2.0 |
|
|
|
|
|
≤ 2.0% |
5 |
5.0 |
|
|
|
|
|
≤ 2.0% |
Observation table for Channel A and B
Eluent Composition |
Hx |
Calculated Bx (%) |
Acceptance Criteria |
20% |
|
|
19.4-20.6 |
40% |
|
|
38.8-41.2 |
60% |
|
|
58.2-61.8 |
80% |
|
|
77.6-82.4 |
100% |
|
|
97.0-103.0 |
Height at 100% composition of B (H100) = __________
Observation table for Channel C and D
Eluent Composition |
Hx |
Calculated Dx (%) |
Acceptance Criteria |
20% |
|
|
19.4-20.6 |
40% |
|
|
38.8-41.2 |
60% |
|
|
58.2-61.8 |
80% |
|
|
77.6-82.4 |
100% |
|
|
97.0-103.0 |
Height at 100% composition of D (H100) = __________
Calculation:
The actually obtained mobile phase composition can be calculated from the Hx values by the following formula.
BX or DX (%) = HX (mm)/H100 (mm)————— x (100)
= _____________________
Where:
X = for the programmed step height (20, 40, 60 and 80).
HX = Height of the respective plateaus in mm of B or D
DX = the actual obtained mobile phase composition.
H100 = Height at 100% composition of B or D.
-
Calibration of Auto Injector:-
Sr No |
Vial Position/ Location |
Observation |
1 |
|
Septum pierced/Not pierced |
2 |
|
Septum pierced/Not pierced |
3 |
|
Septum pierced/Not pierced |
4 |
|
Septum pierced/Not pierced |
5 |
|
Septum pierced/Not pierced |
-
Volume accuracy of injection:-
Instrument/Equipments/Standards used:
Balance No |
|
Balance calibration due date |
|
Injection Volume |
20µl |
Number of Injections |
10 |
Density of Water at 200C (d) |
|
Observation Table:
Initial weight of Vial + Water(W1) in gram |
|
Weight of Vial + Water after 10 Injection(W2) in gram |
|
W = (W1-W2) (in gram) |
|
mL of water injected = W/d in mL |
|
mL of water injected in 10 injection = W/d *1000 µl=A |
|
µl per injection=A/10 |
|
Acceptance Criteria:
The amount of water drawn by the auto injector per injection should be in between 19.0 µl and 21.0µl (i.e., ± 1.0 µl of the set value)
Instrument/Equipments/Standards used:
Balance No |
|
Balance calibration due date |
|
Standard used |
|
Standard weight |
|
Standard Lot number |
|
HPLC Column details |
|
Preparation of Standard Solution (20 ppm Caffeine):
- Weighed ______ mg of Caffeine in 100 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
- Further dilute 10mL of this solution to 100mL with Mobile Phase.
Observation Table
Sr. No |
Injection Volume |
Area |
1 |
5 µl |
|
2 |
10 µl |
|
3 |
20 µl |
|
4 |
50 µl |
|
5 |
100 µl |
|
Correlation coefficient(r2)
(Acceptance criteria r2 ≥ 0.999) |
|
Instrument/Equipments/Standards used:
Balance No |
|
Balance calibration due date |
|
Standard used |
|
Standard weight |
|
Standard Lot number |
|
HPLC Column details |
|
Preparation of Standard Solution (20 ppm Caffeine):
- Weighed ______ mg of Caffeine in 100 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
- Further dilute 10mL of this solution to 100mL with Mobile Phase.
Observation Table
Sr. No. |
Area |
1 |
|
2 |
|
3 |
|
4 |
|
5 |
|
6 |
|
Average Area |
|
RSD (%) |
|
Acceptance Criteria: RSD NMT 2.0 %
Instrument/Equipments/Standards used:
Balance No |
|
Balance calibration due date |
|
Standard used |
|
Standard weight |
|
Standard Lot number |
|
HPLC Column details |
|
Preparation of Standard Solution (500 ppm Caffeine):
- Weighed ______ mg of Caffeine in 10 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
Observation Table
Name |
Area |
% Carryover in Blank |
Blank-1 |
|
NA |
Sample |
|
NA |
Blank-2 |
|
|
Sample |
|
NA |
Blank-3 |
|
|
Sample |
|
NA |
Blank |
|
|
Acceptance Criteria: NMT 0.02 %
(Area response of blank-2) – (Area response of blank-1)
Carryover % = ——————————————————— X 100
Average peak area of caffeine obtained from standard solution
Carryover % = ———————————————————
-
Calibration of Detector:-
Instrument/Equipments/Standards used:
Preparation of Standard Solution (20 ppm Caffeine):
- Weighed ______ mg of Caffeine in 100 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
- Further diluted 10mL of this solution to 100mL with Mobile Phase.
Observation Table for wavelength 205 nm
Wavelength nm |
Response |
Wavelength Maxima/ Maximum response at nm |
Acceptance Criteria |
Area |
Height |
201 |
|
|
|
205nm ± 2 nm |
202 |
|
|
203 |
|
|
204 |
|
|
205 |
|
|
206 |
|
|
207 |
|
|
208 |
|
|
209 |
|
|
Observation Table for wavelength 245 nm
Wavelength nm |
Response |
Wavelength minima/ Minimum response at nm |
Acceptance Criteria |
Area |
Height |
241 |
|
|
|
245nm ± 2 nm |
242 |
|
|
243 |
|
|
244 |
|
|
245 |
|
|
246 |
|
|
247 |
|
|
248 |
|
|
249 |
|
|
Observation Table for wavelength 273 nm
Wavelength nm |
Response |
Wavelength Maxima/ Maximum response at nm |
Acceptance Criteria |
Area |
Height |
269 |
|
|
|
273nm ± 2 nm |
270 |
|
|
271 |
|
|
272 |
|
|
273 |
|
|
274 |
|
|
275 |
|
|
276 |
|
|
277 |
|
|
Instrument/Equipments/Standards used:
Preparation of Standard Solution (0.5mg/mL Caffeine):
Weighed ______ mg of Caffeine in 50 mL volumetric flask, dissolved and diluted up to the mark with Mobile Phase.
- 0.25mg/mL standard Solution:
Diluted 1mL of stock solution to 20mL with mobile phase
- 0.5mg/mL standard Solution:
Diluted 2mL of stock solution to 20mL with mobile phase
- 100mg/mL standard Solution:
Diluted 4mL of stock solution to 20mL with mobile phase
- 125mg/mL standard Solution:
Diluted 5mL of stock solution to 20mL with mobile phase
- 250mg/mL standard Solution:
Diluted 10mL of stock solution to 20mL with mobile phase
- 500mg/mL standard Solution:
Use stock solution
Observation Table:
Theoretical Conc.
(mg/ ml) |
Actual Conc.
(mg/ ml) |
Area of injection 1 |
Area of injection 2 |
Mean Area |
Correlation Coefficient |
0.025 |
|
|
|
|
|
0.050 |
|
|
|
|
0.100 |
|
|
|
|
0.125 |
|
|
|
|
0.250 |
|
|
|
|
0.500 |
|
|
|
|
Acceptance Criteria: Correlation coefficient ≥ 0.999
Acceptance Criteria and results for DAD:
Description |
Acceptance criteria |
Results |
Status Pass/Fail |
Lowest intensity in range 190 nm to 220 nm |
Not less than 2000 cts. |
|
|
Lowest intensity in range 221 nm to 350 nm |
Not less than 5000 cts |
|
|
Lowest intensity in range 351 nm to 500 nm |
Not less than 2000 cts |
|
|
Lowest intensity in range 501 nm to 950 nm |
Not less than 4000 cts |
|
|
Highest intensity in range 190 nm to 350 nm |
Not more than 450000 cts |
|
|
Highest intensity in range 700 nm to 950 nm |
Not more than 300000 cts |
|
|
Highest intensity for D2 alpha line |
Not more than 1200000 cts |
|
|
Acceptance Criteria and results for VWD:
Description |
Acceptance criteria |
Results |
Status Pass/Fail |
Highest intensity |
More than 10000 cts |
|
|
Lowest intensity |
More than 200 cts |
|
|
Average intensity |
More than 5000 cts |
|
|
-
Calibration of column Compartment:-
Observation Table for Right side Compartment
Set Temperature (in °C) |
Displayed Temperature on instrument
( in° C) |
Observed Temperature on thermometer
( in° C) |
Difference between set temp and observed temp on thermometer ( in° C) |
25.0 |
|
|
|
40.0 |
|
|
|
60.0 |
|
|
|
80.0 |
|
|
|
Observation Table for Left side Compartment
Set Temperature (in °C) |
Displayed Temperature on instrument
( in° C) |
Observed Temperature on thermometer
( in° C) |
Difference between set temp and observed temp on thermometer ( in° C) |
25.0 |
|
|
|
40.0 |
|
|
|
60.0 |
|
|
|
80.0 |
|
|
|
Acceptance Criteria:
All the observed thermometer temperature values are within ± 2°C of set value.
-
Calibration of Autosampler Temperature:-
Observation Table
Set Temperature (in °C) |
Displayed Temperature on instrument
( in° C) |
Observed Temperature on thermometer
( in° C) |
Difference between set temp and observed temp on thermometer ( in° C) |
25.0 |
|
|
|
20.0 |
|
|
|
10.0 |
|
|
|
4.0 |
|
|
|
Acceptance Criteria:
All the observed thermometer temperature values are within ± 2°C of set value.
Remark: System complies/does not comply with respect to above test.
Performed By: Checked By:
Date: Date: