SOP Title-Procedure for Fumigation in Microbiology
- OBJECTIVE
To lay down the procedure for fumigation in microbiology area.
- SCOPE
This SOP is applicable for the procedure for fumigation in Microbiology Lab.
- RESPONSIBILITY
- QC Executive and above shall be responsible for implementation of this procedure.
- QC Head / designee shall be responsible for compliance of this procedure.
- PROCEDURE :
- Precaution
- Ensure that air handling unit is switched off before fumigation activity is carried out.
- Check and ensure that any material should not be opened.
- Wear nose mask, cap and surgical hand gloves during fumigation.
- Fumigation
- Prepare the required quantity of 3 % gramicid solution for fumigation.
- Pour the prepared solution into molded tank of fumigator.
- Place the fumigator in the respective area where the fumigation activity is carried out.
- Adjust the ejector to ensure that the fumigating solution covers the whole area.
- Connect the fumigator with electric supply and switch ON the main supply.
- Leave the room and close the door of the area immediately.
- Fumigation of microbiological shall be carried out for 20-30 minutes.
- Switch “OFF” the fumigator after fumigation.
- Area shall be locked and labeled as ‘FUMIGATED AREA-DO NOT ENTER”.
- Fumigated area shall be open after minimum 12 hours.
- AHU shall be started for 10-15 minutes before entering in the area for release the fumes.
- Fumigation record shall be maintained in annexure.
- Precaution
- ABBREVIATION(s)
Abbreviation | Full Description |
SOP | Standard Operating Procedure |
AHU | Air Handling Unit |
% | Percent |
*Note – Ready to use SOP available in “DOWNLOAD” Section.