SOP Title: Analytical Method Validation, Analytical Method Verification and Method Transfer
- Objective:
To lay down the procedure for analytical method validation, analytical method verification and method transfer in Quality Control Department.
- Scope:
This SOP is applicable for analytical method validation, analytical method verification and method transfer in Quality Control Department.
- Responsibility:
- Quality Control: To prepare and review the SOP. To follow the procedures laid down for Validation of Analysts in Quality Control Department as per this SOP.
- Quality Assurance Department: To review and approve the SOP and Annexures.
- Accountability:
Head – Quality Control / Head – Quality Assurance
- Procedure:
- Operational Procedure:
5.1.1 | The analytical method validation protocol shall be prepared for Method Validation and Method Transfer. | |||||||||||
5.1.2 | The analytical method validation protocol shall contain the following parameter;
a. Analytical method validation protocol cover page b. Protocol approval c. Revision history d. Objective e. Scope f. Responsibility g. Methods reference number
|
|||||||||||
5.1.3 | A. Specificity
B. Forced degradation (if applicable) C. Precision · Repeatability · Intermediate precision · Reproducibility D. Linearity E. Limit of detection and limit of quantitation in case of related substances and cleaning samples F. Accuracy G. Robustness H. Solution stability I. Filter suitability (if applicable) J. Swab recovery in case method validated for cleaning samples |
|||||||||||
5.1.4 | Above all parameters are not limited during the analytical method validation. These parameters can be include and exclude on the basis of specific regulatory requirements/marketing authorization holders/customers. | |||||||||||
5.1.5 | Recording of raw data during analytical method validation shall be recorded in worksheet issued by QA. | |||||||||||
5.1.6 | After execution of analytical method validation as per their respective protocol, analytical method validation report shall be prepared. | |||||||||||
Analytical method validation | ||||||||||||
5.1.7 | The analytical method validation report shall contain the following parameter; | |||||||||||
5.1.8 | Analytical method validation report cover page
a. Report approval b. Revision history c. Objective d. Scope e. Responsibility f. Methods reference g. Validation summary report h. Procedure i. Specificity j. Forced degradation if applicable. k. Precision · Repeatability · Intermediate precision · Reproducibility (If applicable) l. Linearity m. Range |
|||||||||||
5.1.9 | Limit of detection (LOD) and limit of quantitation (LOQ) in case of related substances and cleaning validation samples.
i. Accuracy ii. Robustness iii. Solution stability iv. Filter compatibility v. Swab recovery (If applicable) vi. Deviation if any vii. Conclusion |
|||||||||||
5.1.10 | Complete analytical method shall be included in the analytical method validation protocol and report as per respective standard test procedure. | |||||||||||
5.1.11 | An approval of analytical method validation protocol / analytical method validation report shall be taken from marketing authorization holder/customer, in case of product manufactured for contract giver. | |||||||||||
5.1.12 | Procedure for analytical method validation shall be as follows; |
|
||||||||||
5.1.13 | Specificity:
Ability to asses unequivocally the analyte in the presence of components which may be expected to be present (Impurities, degradants) |
|||||||||||
5.1.14 | For assay and impurity methods the samples shall contain materials which are potentially present during routine analysis and may interfere with the result, e.g. analysis sample containing potential interferences, e.g. impurities, Excipients. | |||||||||||
5.1.15 | Sample shall prepared by spiking drug substance / drug product with potentially interfering material. The acceptance criteria shall be; No interference observed for response due to analyte or impurities or interest and no peaks interfering with analyte peak or the peaks due to the impurities of interest observed. Peak purity shall not be less than 0.995. | |||||||||||
5.1.16 | When the impurities are available, the specificity of an assay method is determined by comparing the impurities, degradation products or placebo ingredients with those obtained from an analysis of sample without the impurities, degradation products or placebo ingredients. | |||||||||||
5.1.17 | For the impurities test, the determination shall be established by spiking drug substances or drug product with appropriate levels of impurities and demonstration the separation of these impurities individually or from other components in the sample matrix. | |||||||||||
5.1.18 | In case impurities are not available. Specificity shall be demonstrated by comparing test results with the blank and placebo ingredients. | |||||||||||
5.1.19 | Forced degradation:
Forced degradation shall be done in presence of excipients, firstly performed during the pre-formulation stage to assist in the selection of the most formidable compounds and excipients. This shall lead to the development of more suitable formulation, packaging and change in storage and manufacturing conditions as the optimal formulation is defined to be used. Forced degradation studies are also in order to demonstrate specificity during the validation of stability indicating methods. These studies are usually performed at conditions exceeding that of accelerated storage condition. Forced degradation studies shall provide information to degradation pathway and degradation products that could form during storage of the drug product. The main goal of forced degradation studies is to effectively produce samples containing representative and realistic degradation products. These degradation products shall be assessed whether they are a related to the excipients or due to drug substance- excipients interaction under certain forced degradation conditions. A delicate balance of efficiency and severity / duration of stress conditions are needed. Overstressing shall destroy relevant compounds or generate irrelevant compounds. Under stressing may fail to generate important degradation products. The extent of targeted degradation shall be approximately anywhere from 5% to 20%. In some cases the degradation can be difficult to achieve 5%, in such cases statement shall be made in the validation report that the product quality / efficacy shall not be affected for such particular stress condition. The assessment of peak purity using diode array shall be employed. The stress testing shall be performed as per the below table;
|
|||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.20 | Precision:
Precision is the closeness of agreement (degree of scatter) between a series of measurements obtained from multiple sampling of the same homogeneous sample. There are following aspects to achieve for precision. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.21 | Repeatability:
Repeatability is a measure of precision under the same operating conditions over a short interval of time that is, under normal operating conditions of analytical method with the same equipment. Repeatability to be accessed using a minimum of nine determinations covering the specified range for the procedure (e.g. three concentrations / three replicates as in the accuracy experiment) of using a minimum six determination at 100% of the test concentration. Standard deviation, relative standard deviation and confidence interval to be reported. The assay value of independent analysis of sample preparation to final test result. For the related substances, residual solvents precision shall be established at 100% specification level of know impurities / organic solvents and also to be established at the limit of quantification level. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.22 | Intermediate precision:
Intermediate precision is defined as the variation within the same laboratory. The extent to which intermediate precision needs to be established depends on the circumstances under which the procedure is intended to be used. Typical parameters that are investigated include day to day variation, analyst variation and equipment variation. Depending upon the extent of the study, the use of experimental design is encouraged. Experimental design shall be minimize the number of experiments that need to be performed. It is important that exemption from doing intermediate precision when reproducibility is proven. It shall be expected that the intermediate precision shall show variability that is in same range or less than repeatability variation. The standard deviation, relative standard deviation (coefficient variation) and confidence interval to be established of data. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.23 | Reproducibility:
Reproducibility measures the precision between laboratories. This parameter shall be considered in the standardization of an analytical method (e.g. inclusion of procedures in pharmacopoeias and method transfer between different laboratories). To validate this characteristic, similar study to be performed at different laboratories using the same homogeneous sample lot and the same experimental design. In case of method transfer between two laboratories, different approaches may be taken to achieve the successful transfer of the procedure. The approach is method transfer from the originating laboratory to the receiving laboratory. The originating laboratory is defined as the laboratory that has developed and validated analytical method. The receiving laboratory is defined as laboratory to which the analytical method to be transferred and that will participate in the method transfer studies. In the method transfer, it is recommended that protocol be initiated with details of the experiments to be performed and acceptance criteria (in term of the difference between the means of the two laboratories) for passing the method transfer. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.24 | Linearity:
Linearity of an analytical procedure as the ability (within the given range) to obtain test results of variable data (e.g. absorbance and area under the curve) which are directly proportional to the concentration (amount of analyte) in the sample. The data variables that shall be used for quantitation of the analyte are the peak area or ration of peak area of analyte to the internal standard peak. The working sample concentration and samples tested for accuracy shall be in the linear range. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.25 | There are two general approaches for determining the linearity of the method. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.26 | The first approach is to weigh different amounts of standard directly to prepare linearity solutions at different concentrations. However, it is not suitable to prepare solution at very low concentration, as the weighing error shall be relatively high. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.27 | Another approach is to prepare a stock solution of high concentration. Linearity is then demonstrated directly by dilution of standard stock solution. This is more popular and the recommended approach.
Linearity is best evaluated by visual inspection of a plot of signals as a function of analyte concentration. Subsequently, the variable data are generally used to calculate a regression line by the least square method. At least five concentration levels shall be used. Under normal circumstances, linearity shall be with coefficient of determination (r2) of 0.995 the slope, residual sum of squares, and y intercept shall also be reported. The slope of the regression line shall provide an idea of the sensitivity of the regression. The y intercept shall provide an estimate of the variability of the method. The ratios per cent of the y- intercept with the variable data at nominal concentration are used to estimate the method variability. For determination of assay of drug substances of a drug product, the usual range of linearity shall be ±20% of the target or nominal concentration. For the determination of content uniformity, it shall be ±30% of the target of nominal concentration. For determination of related substances or residual solvents it shall be from the reporting level to 20% ahead of the target or nominal concentration. For dissolution testing: ±20% over the specified range. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.28 | Range:
The range of analytical procedure is the interval between the upper and lower concentration of analyte in the sample for which it has been demonstrated that the analytical procedure shall a suitable level of precision, accuracy, and linearity. The range shall normally expressed in the same units as test results (e.g. percent, parts per million) obtained by the analytical procedure. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.29 | For the assay of drug substance or finished drug product, it is normally recommended to have a range of 80 – 120% of the nominal concentration. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
For content uniformity, a normal range shall cover 70 – 130% of the nominal concentration, unless a wider and more appropriate range is justified.
For dissolution testing, a normal range is ± 20% over the specified range. In case the acceptance criteria for a controlled – release product covers a region from 20% after1 h, and up to 90% after 24 h, the validated range would be 0 – 110% of the label claim. In this case, the lowest appropriate quantifiable concentration of analyte will be used as the lowest limit as 0% is not appropriate. For the related substances / residual solvents of drug substance or finished drug product, it is normally recommended to have a range of limit of quantitation – 120% of the nominal concentration. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.30 | Limit of Detection (LOD):
The Limit of Detection (LOD) is the lowest amount of analyte in a sample that shall be detected but not necessarily quantitated under the stated experimental conditions. The detection shall usually expressed as the concentration of the analyte in the sample, e.g. percentage, parts per million (ppm) or parts per billion (ppb). There are several approaches to establish the limit of detection. One approach to establish detection limit shall be determined by the analysis of a series of sample with known concentrations and establishing the minimum level at which analyte shall be reliably detected. Presentation of relevant chromatograms or other relevant data shall be sufficient for justification of the detection limit. For instrumental procedures that exhibit background noise, it shall compare measured signals from samples with known concentrations of analyte with those of the blank samples. The minimum concentration at which the analyte shall reliably be detected, shall establish using an acceptable signal to noise ratio of 2:1 or 3:1. Presentation of relevant chromatograms shall be sufficient for justification of detection limit. Another approach to estimate the detection limit from the standard deviation of the response and the slope of the calibration curve. The standard deviation shall be determined either from the standard deviation of multiple blank samples or from the standard deviation of the y intercepts of the regression lime done in the range of the detection limit. This estimate shall need to be subsequently validated by the independent analysis of a suitable number of samples near or at the detection limit; |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Where, σ = the standard deviation of the response S = the slope of the calibration curve
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.31 | Limit of Quantitation (LOQ):
The Limit of Quantitation (LOQ) is a characteristic of quantitative assays for low levels of compounds in the sample matrices, such as impurities in bulk drug substances and degradation products in finished pharmaceuticals. LOQ shall be defined as the concentration of related substances in the sample that shall give a signal to noise ratio of 10:1; the LOQ of a method is affected by both the detector sensitivity and the accuracy of sample preparation at the low concentration of the impurities. In practice LOQ shall be lower than the corresponding report limit. Generally, three approaches are to establish the quantitation limit. The first approach is to evaluate it by visual evaluation and shall be used for non instrumental methods and instrumental methods. Quantitation limit is determined by analysis of samples with known concentrations of analyte and by establishing the minimum level at which the analyte shall be quantitated with acceptable accuracy and precision. The second approach determines the signal to noise ratio by comparing measured signals from samples with known low concentration at which the analyte shall be reliable quantified at the signal to noise ration 10:1. The third approach estimates LOQ by equation as per limit of detection;
Where, σ = the standard deviation of the response S = the slope of the calibration curve The slope shall be established from calibration curve of the analyte. The value of the σ shall be established by · Calculating the standard deviation of the responses obtained from the measurement of the analytical background response of an appropriate number of blank samples or · Calculating the residual standard deviation of the regression line from the calibration curve using samples containing the analyte in the range of LOQ. Whatever approach is applied, the LOQ shall be subsequently validated by the analysis of a suitable number of samples prepared at the LOQ and determining the precision and accuracy at this level. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.32 | Accuracy:
Accuracy of an analytical procedure as the closeness of agreement between the values that are accepted either as conventional true values or an accepted references value and the value found. For the drug substances accuracy is defined by the application of the analytical procedure to an analyte of known purity (e.g. reference standard). For the drug product, accuracy shall be determined by application of the analytical procedure to synthetic mixtures of the drug product components to which amounts of analyte have been added within the range of the procedure. It is recommended to assess minimum of nine determinations over a minimum three concentration levels covering the specified range (e.g. three concentration / three replicates). Accuracy shall be reported as percent recovery by assay (using the proposed analytical procedure) of known added amount of analyte in the sample of as the difference between the mean and the accepted true value together with the confidence intervals. The range for the accuracy limit shall be within the linear range. Typically accuracy of the recovery of the drug substance is expected to be about 99 – 101%. Typical accuracy of drug product is expected to be 98 – 102%. For the dissolution of drug product accuracy of recovery is expected to be 98 – 102%. Values of accuracy of recovery data beyond this range need to be investigated as appropriate. For the related substances / residual solvents recovery shall be performed from the quantitation limit by spiking the known concentration of known impurities / residual solvents in analyte and the accuracy of recovery can be up to 85 – 115%. For Cleaning validation recovery shall be performed from the quantitation limit to minimum 120% by spiking the known concentration of known analyte and the accuracy of recovery should not be less than 70%. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.33 | Robustness:
Robustness of an analytical procedure is a measure of the analytical method to remain unaffected by small but deliberate variations in method parameter and provides an indication of its reliability during normal usage. Common method parameters that can affect the analytical procedure shall be considered based on the analytical technique; . Sample preparation . Extraction time . Sample solvent (pH ±0.2, percent organic ± 5% absolute) |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
· High performance liquid chromatography (HPLC) condition / spectrophotometric methods
Mobile phase composition (pH ±0.05, percent organic ± 2% absolute) Column used (equivalent columns, lots and / or suppliers) Temperature (± 5º C) Flow rate (± 10 %) · Gas chromatography (GC) conditions Column used (equivalent columns, lots and / or suppliers) Temperature (± 10 %) Flow rate (± 10 %) When the results are affected by some critical experimental parameters, a precautionary statement shall be included in the analytical procedure to ensure that this parameter is tightly controlled between experiments. For example: in case percent ion pairing of mobile phase affects the results significantly, the analytical procedure shall explicitly be written with a precautionary statement for aqueous component. Other robustness considering for ruggedness of the analytical procedure during validation include the following; · Sample extraction: mechanical shaking preferred over sonication as the latter is affected by a number of factors, for · example, time for sonication. · Dilution of sample and solvent, minimize the number of dilution steps to reduce introduction of error. Dilution solvent shall be as similar to mobile phase as possible. |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.34 | Solution stability:
· Stability of Standard and Sample Solution shall be performed at various time intervals and at room temperature. · Acceptance criteria shall be as follows : For assay and dissolution, the relative standard deviation for peak area as obtained from standard solution and test solution shall not be more than 2.0%. For related substance and residual solvents, the relative standard deviation for peak area as obtained from standard solution and test solution shall not be more than 10.0 % at various time intervals |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.35 | Filter compatibility:
Filter compatibility shall be performed with different filter against the centrifuged solution. Acceptance criteria for filter compatibility shall be as follows: The similarity factor for centrifuged standard and filtered standard solution shall be between 0.98 and 1.02. The similarity factor for centrifuged sample solution and filtered sample solution shall be between 0.98 and 1.02. |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.36 | Following parameters to be established during the analytical method validation; | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
5.1.37 | Method Verification:
Verification consists of assessing selected analytical validation characteristics described earlier to generate appropriate relevant data rather than repeating the validation process for commercial products. The method verification shall be done for the pharmacopeial methods (compendial methods) such as titrations, chromatographic procedures (related compounds, assay, dissolution and limit tests), and spectroscopic tests. However, general tests (water, heavy metals, residue on ignition) do not typically require verification. |
|||||||||||||||||||||||||
5.1.38 | The analytical method verification protocol shall be prepared. | |||||||||||||||||||||||||
5.1.39 | The analytical method verification protocol shall contain the following parameter;
Analytical method validation protocol cover page Protocol approval Revision history Objective Scope Responsibility Methods reference number Procedure Precision · Repeatability · Intermediate precision · Reproducibility |
|||||||||||||||||||||||||
5.1.40 | Specificity :
|
|||||||||||||||||||||||||
5.1.41 | Method Revalidation:
The analytical method shall be revalidation under following circumstances; however these circumstances are not limited; · In case new impurity found that makes deficient in its specificity, this method needs to be modified and revalidated. · Changes in the excipients composition can change the product impurity profile. · Changes in equipment or supplier of critical supplies of the API (Active Pharmaceutical Ingredients) or final drug product shall have the potential to change their degradation profile and shall require the method to be redeveloped and revalidated. |
|||||||||||||||||||||||||
- Definitions /Abbreviations:
- Definitions:
- Abbreviations:
Abbreviation | Expansion |
e.g. | Example |
No. | Number |
°C | Degree Centigrade |
RH | Relative Humidity |
% | Percent |
N | Normal |
ppm | Parts per million |
ppb | Parts per billion |
USP | United State of Pharmacopoeia |
- Reference:
References | Description |
Q2 (R1) | Validation of analytical procedures : Text and Methodology. |
USP <1224> | Transfer of analytical procedures |
USP <1225> | Validation of compendial procedures |
USP <1226> | Verification of compendial procedures |
*Note – Ready to use SOP available in “DOWNLOAD” Section.