Categories
Microbiology SOP

SOP Title: Procedure for Validation of Laminar Air Flow

SOP Title: Procedure for Validation of Laminar Air Flow

  1. Objective:

To lay down the procedure for Validation of Laminar Air Flow / Reverse Laminar Air Flow.

  1. Scope

This procedure is applicable for Validation of LAF in the Microbiology, Sampling Booth RLAF and Dispensing Booth RLAF.

  1. Responsibility:

3.1   Quality Control Department: To prepare and review the SOP. To follow the procedures for  Validation of LAF as per this SOP.

  • Quality Assurance Department: To review and approve the SOP and Annexure.
  1. Accountability:

Head Quality Control Department, Head Quality Assurance Department.

Procedure:

     
   5.1 Safety Precautions/Instructions:
5.1.1 Ensure that Laminar Air Flow Unit is properly installed in a clean environment.  
5.1.2 Ensure that pre filters of Laminar Air Flow Unit are clean.  
5.1.3 Before Starting the activity ensure that LAF is properly Cleaned.  
   5.2 Procedure for Validation:
5.2.1 Validation of LAF is carried out by performing following tests.

1. HEPA Filter Integrity by PAO Test.

2. Air flow Velocity: Unidirectional Airflow.

3. Non-Viable Particle Count Test.

 
5.2.2  Validation study of LAF is applicable to LAF in Microbiology, Sampling  Booth RLAF and Dispensing Booth RLAF.  
   5.3 1. HEPA Filter Integrity by PAO Test:
5.3.1 LAF consists of two filters and are named as F1(Left side) and F2(Right side).  
5.3.1 Arrange single phase power supply as well as air connection with 20psi (1.5kg) pressure.  
5.3.2 Switch ON Aerosol Generator System, before starting the measurements and Connect the Spike Guard to Photometer.  
5.3.3 Start the air to generate the aerosol particles of PAO at 20 psi pressure by using Aerosol Generator System.  
5.3.4 Direct the generator face to the return air point or fresh air intake of the LAF.  
5.3.5 Connect one end of the tube to filter face upstream port and another end to Photometer in Downstream port.  
5.3.6 Switch ON the photometer and set it to zero.  
5.3.7 Put the photometer to upstream mode and visually note the reading on display, the reading should be in between 20 mg/Ltr to 80 mg/Ltr.  
5.3.8 If it is between the above range then set the value to 100%, wait until the photometer displays 100% upstream concentration.  
5.3.9 Switch ‘ON’ the photometer, enter the clear mode and enter zero. Wait until photometer display shows ‘0’ (zero).  
5.3.10 Change photometer mode to downstream mode and scan the entire filter by holding the probe approximately one inch away from the filter face.  
5.3.11 Scan the corners of HEPA filter for checking any leakages.  
5.3.12 If the reading on the photometer is more than 0.01%, rectify the leakages by tightening the locking bolts.  
5.3.13 Repaired patches on filters should not exceed maximum of 5% of the total filter face area and the maximum width x length of each patch should not be more than 1.5 inches.  
5.3.14 Total number of patches should not exceed 5 numbers / filters.  
5.3.15 Note down the final readings after rectification.  
5.3.16 If the reading on the Photometer is more than 0.01% or total number of patches exceed 5 numbers per filter, then HEPA Filter Integrity Test fails.  
5.3.17 Replace with new HEPA Filter and carry out PAO Test again.  
5.3.18 Acceptance Criteria: During scanning, percentage of the PAO penetration shown by photometer should be less than 0.01% throughout the filter.  
   5.4 2. Air Flow Velocity: Unidirectional Airflow
5.4.1 Switch ‘ON’ Anemometer for measuring Air Velocity before starting the measurement. Use Vane type anemometer.  
5.4.2 1        Filter face Air Velocity should be measured at distance of about 150 mm.  
5.4.3 Switch ‘ON’ the LAF air flow,  place the Anemometer in front of the filter surface at a distance of about 150 mm. Check the air velocity readings at the four corners and centre. Note readings as 1, 2, 3, 4, and 5 on annexure no. 01.

Refer below figure.

       1                                                                         4

                                   5

2                                                                                     3

 
5.4.4 Measuring time at each position should be at least 10 seconds. Record the average, maximum and minimum values. Note down the readings in annexure.  
5.4.5 Acceptance Criteria: The measured average velocity for each filter shall be within 0.36m/s to 0.54 m/s as per ISO-14644.  
  5.5 3. Non- Viable Particle Count test.:
5.5.1 The test method specifies the measurement of air borne particle concentration with size distributions between 0.5µm and 5.0µm.  
5.5.2 Measurements can be made in any of three states as-built, at-rest and operational.  
5.5.3 LAF Comes under ISO CLASS 5 Classifications.

At rest the particles should not exceed 3520 (0.5µm) size and 29 (5 µm) size for compliance of LAF for Non- Viable Particle Count test.

 
5.5.4 Switch ‘ON’ the LAF for 2 hours before starting the measurements.  
5.5.5 Connect the Spike Guard to particle counter and set the Particle counter at center position or as per the the working level of LAF.  
5.5.6 Position the sampling probe vertically upward.  
5.5.7 Switch ‘ON’ the Particle Counter and adjust the Particle Counter for ‘0’ display.  
5.5. 8 Start taking readings of LAF and take print of the reading. Take at least 3 readings from same position in LAF. Refer ISO-14644 for the test.  
5.5.9 Thermal paper print should be pasted on a blank page and should be Photocopied immediately.  
5.5.10 Record validation data on Annexure.  
5.5.11 Acceptance Acceptance Criteria: The LAF should meet the criteria as per ISO Class 5 classifications i.e. NMT 3520 particles/m3, for particles size equal to or greater than 0.5 µm and NMT 29 particles/m3 for particles size equal to or greater than 5 µm.  
  Frequency – Yearly by External agency
5.5.12       After the test is completed, the LAF and the surrounding areas shall be cleaned properly.  
5.5.13        If the validation study fails, file an deviation, rectify the problem and again validate the LAF.  
     

 

  1. Definitions / Abbreviations :
  • Definitions :

 

  • Abbreviations :
Abbreviation Expansion
ISO International Standard Operating
RLAF Reverse Laminar Air Flow
LAF Laminar Air Flow
AHU Air Handling Unit
psi per square inch
HEPA High efficiency particulate air
PAO Poly Alfa  Olefin
mg milligram
% percent
lts litres
IPA Iso Propyl Alcohol
ft feet
avg average
μm micrometer

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 

Leave a Reply

Your email address will not be published. Required fields are marked *

error: Content is protected !!