SOP Title-Operation and Cleaning of Antibiotic Zone Reader

SOP Title-Operation and Cleaning of Antibiotic Zone Reader

 

  • OBJECTIVE

To lay down the procedure for operation and cleaning of antibiotic zone reader.

  • SCOPE

This SOP is applicable for operation and cleaning of antibiotic zone reader in Microbiology Lab.

  • RESPONSIBILITY
  • QC Executive and above shall be responsible for Implementation of this procedure.
  • QC Head / designee shall be responsible for compliance of this procedure.
  • PROCEDURE :
  • Precautions
  • Check the cleaning status of antibiotic zone reader.
  • Check the calibration status of antibiotic zone reader.
  • Operation of Antibiotic Zone Reader :
  • Switch ‘ON’ the mains switch.
  • Turn the left side drum to its maximum clockwise position, so that zero marking on the scale of the right side drum coin­cides with zero marking on the vernier scale.
  • Keep the petri dish on the aluminium plate and rotate it manu­ally; so that one end of the zone image touches the mark line on the prism.
  • Now turn the left side drum slowly, in anti-clockwise posi­tion, till the other end of the zone image touches the mark line on the prism.
  • Note the reading directly on the right side scale.
  • The scale is marked from 0 to 35 mm.
  • The side scale having each division equal to 0.2 mm.
  • Remove the petri dish and switch ‘Off’ the mains.
  • record shall be maintained as per instrument usage log book.
  • Cleaning:
  • After use, Instrument shall be cleaned with 70% IPA by using tissue paper/clean lint free cloth.

 

  • ABBREVIATION(s)
Abbreviation Full Description
SOP Standard Operating Procedure
IPA Isopropyl Alcohol

 

*Note – Ready to use SOP available in “DOWNLOAD” Section.

 