{"id":812,"date":"2025-06-14T15:32:59","date_gmt":"2025-06-14T15:32:59","guid":{"rendered":"https:\/\/pharmacoguide.com\/?p=812"},"modified":"2025-06-14T15:32:59","modified_gmt":"2025-06-14T15:32:59","slug":"sop-title-operation-and-calibration-of-micrometer-screw-gauge","status":"publish","type":"post","link":"https:\/\/pharmacoguide.com\/?p=812","title":{"rendered":"SOP Title: Operation and Calibration of Micrometer Screw Gauge"},"content":{"rendered":"<p><strong><u>SOP Title:<\/u><\/strong> <strong><u>Operation and Calibration of Micrometer Screw Gauge<\/u><\/strong><\/p>\n<ol>\n<li><strong>Objective:<\/strong><\/li>\n<\/ol>\n<p>To lay down the procedure for Operation and Calibration of Micrometer Screw Gauge.<\/p>\n<ol start=\"2\">\n<li><strong>Scope:<\/strong><\/li>\n<\/ol>\n<p>This SOP is applicable for Operation, Cleaning and Calibration of Micrometer Screw Gauge located in Quality Control Laboratory.<\/p>\n<p><strong>\u00a0<\/strong><\/p>\n<ol start=\"3\">\n<li><strong>Responsibility:<\/strong><\/li>\n<\/ol>\n<ul>\n<li>Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation and Calibration of Micrometer Screw Gauge as per this SOP.<\/li>\n<li>Quality Assurance Department: To review and approve the SOP and Annexures.<\/li>\n<\/ul>\n<p><strong>\u00a0<\/strong><\/p>\n<ol start=\"4\">\n<li><strong>Accountability:<\/strong><\/li>\n<\/ol>\n<p>Head Quality Control Department, Head Quality Assurance Department.<\/p>\n<ol start=\"5\">\n<li><strong>Procedure:<\/strong><\/li>\n<\/ol>\n<p><strong>\u00a0<\/strong><\/p>\n<table width=\"692\">\n<tbody>\n<tr>\n<td width=\"67\"><strong>\u00a0\u00a0\u00a0 5.1<\/strong><\/td>\n<td colspan=\"2\" width=\"625\"><strong>Safety Precautions\/Instructions:<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.1.1<\/td>\n<td width=\"607\">Wear personal protective equipments such as hand gloves during operation of micrometer screw gauge.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\"><strong>\u00a0\u00a0\u00a0 5.2<\/strong><\/td>\n<td colspan=\"2\" width=\"625\"><strong>Operation:<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.1<\/td>\n<td width=\"607\">Range of the instrument is 0 to 25 mm &amp; least count is 0.001mm.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.2<\/td>\n<td width=\"607\">There are three scale readings on micrometer screw gauge.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.3<\/td>\n<td width=\"607\">Main scale range is 0 to 25 mm, circular scale range is 0 to 50 and coincide scale range is 0 to 10.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.4<\/td>\n<td width=\"607\">Loosen the small screw provided by pushing it right side.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.5<\/td>\n<td width=\"607\">First adjust the circular scale to the 0 reading.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.6<\/td>\n<td width=\"607\">0 on the main scale should coincide with the 0 of circular scale reading.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.7<\/td>\n<td width=\"607\">Then carefully insert the sample into the micrometer screw gauge by rotating the circular scale knob to hold firmly.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.8<\/td>\n<td width=\"607\">First note down the reading on the main scale.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.9<\/td>\n<td width=\"607\">Then note down the reading on circular scale which is coinciding with the division on the coincide scale.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.10<\/td>\n<td width=\"607\">Calculate the thickness of the sample using the formula:<\/p>\n<p>Main scale reading + (circular scale x least count)<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.11<\/td>\n<td width=\"607\">Example:<\/p>\n<p>If main scale reading is 1 mm and circular scale reading coinciding is 44 then by applying formula, thickness of the sample is,<\/p>\n<p>1 + (44 x 0.001) \u00a0= 1.044 mm.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.12<\/td>\n<td width=\"607\">Remove the sample and bring the instrument in its initial position.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\"><strong>5.3<\/strong><\/td>\n<td colspan=\"2\" width=\"625\"><strong>Calibration Procedure:<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"67\">&nbsp;<\/td>\n<td colspan=\"2\" width=\"625\"><strong>Frequency: <\/strong><strong>\u00a0<\/strong>Once in a year by external agency.<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.3.1<\/td>\n<td width=\"607\">Calibration is carried out by external agency as per SOP \u201cCalibration of Instruments and Equipments\u201d.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.3.2<\/td>\n<td width=\"607\">Refer Annexure for calibration test to be performed and acceptance criteria.<\/p>\n<table width=\"100%\">\n<tbody>\n<tr>\n<td width=\"10%\"><strong>Sr. No.<\/strong><\/td>\n<td width=\"21%\"><strong>Tests<\/strong><\/td>\n<td width=\"31%\"><strong>Set value<\/strong><\/td>\n<td width=\"37%\"><strong>Acceptance Criteria (mm)<\/strong><\/td>\n<\/tr>\n<tr>\n<td rowspan=\"10\" width=\"10%\">1.<\/td>\n<td rowspan=\"10\" width=\"21%\">Gauge size<\/td>\n<td width=\"31%\">2.5 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<tr>\n<td width=\"31%\">5.1 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<tr>\n<td width=\"31%\">7.7 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<tr>\n<td width=\"31%\">10.3 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<tr>\n<td width=\"31%\">12.9 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<tr>\n<td width=\"31%\">15.0 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<tr>\n<td width=\"31%\">17.6 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<tr>\n<td width=\"31%\">20.2 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<tr>\n<td width=\"31%\">22.8 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<tr>\n<td width=\"31%\">25.0 mm<\/td>\n<td width=\"37%\">\u00b1 0.004<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.3.3<\/td>\n<td width=\"607\">Generate the A. R. No. on the \u201cInstrument\/Equipment Calibration Register\u201d as per SOP and give requisition to QA department for calibration formats of instruments\/equipments on the due date of calibration.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.3.4<\/td>\n<td width=\"607\">For issuance of calibration report refer SOP \u201cControl of documents, Preparation, Approval, Issuance and Maintenance.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.3.5<\/td>\n<td width=\"607\">Update \u201cAnnual Calibration Schedule\u201d and \u201cMonthly Calibration Schedule\u201d as per SOP No \u201cCalibration of Instruments and Equipments\u201d.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.3.6<\/td>\n<td width=\"607\">If any breakdown occurs, follow procedure as per SOP \u201cCalibration of Instruments and Equipments\u201d.<\/td>\n<td width=\"19\">&nbsp;<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p><strong>\u00a0<\/strong><\/p>\n<p><strong>\u00a0<\/strong><\/p>\n<ol start=\"6\">\n<li><strong>Definitions <\/strong><strong>\/ Abbreviations:<\/strong><\/li>\n<\/ol>\n<ul>\n<li><strong>Definitions:<\/strong><\/li>\n<li>Breakdown: Any activity leading to operation of instrument\/equipment other than the set parameters or unusual sound or vibration observed in the instrument\/equipment.<\/li>\n<li>Calibration: Tests performed to ensure that the instrument\/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.<\/li>\n<\/ul>\n<p><strong>\u00a0<\/strong><\/p>\n<ul>\n<li><strong>Abbreviations:<\/strong><\/li>\n<\/ul>\n<table>\n<tbody>\n<tr>\n<td width=\"116\"><strong>Abbreviation<\/strong><\/td>\n<td width=\"276\"><strong>Expansion<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"116\">SOP<\/td>\n<td width=\"276\">Standard Operating Procedure<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">QC<\/td>\n<td width=\"276\">Quality Control<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">QA<\/td>\n<td width=\"276\">Quality Assurance<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">A. R. No<\/td>\n<td width=\"276\">Analytical Reference Number<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">No.<\/td>\n<td width=\"276\">Number<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">mm<\/td>\n<td width=\"276\">millimetre<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">PM<\/td>\n<td width=\"276\">Preventive Maintenance<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p><strong>\u00a0<\/strong><\/p>\n<p><strong><em><u>*Note &#8211; Ready to use SOP available in \u201cDOWNLOAD\u201d Section.<\/u><\/em><\/strong><\/p>\n<p><strong>\u00a0<\/strong><\/p>\n","protected":false},"excerpt":{"rendered":"<p>SOP Title: Operation and Calibration of Micrometer Screw Gauge Objective: To lay down the procedure for Operation and Calibration of Micrometer Screw Gauge. Scope: This SOP is applicable for Operation, Cleaning and Calibration of Micrometer Screw Gauge located in Quality Control Laboratory. \u00a0 Responsibility: Quality Control: To prepare and review the SOP. To follow the [&hellip;]<\/p>\n","protected":false},"author":1,"featured_media":0,"comment_status":"open","ping_status":"closed","sticky":false,"template":"templates\/template-full-width.php","format":"standard","meta":{"footnotes":""},"categories":[4],"tags":[],"class_list":["post-812","post","type-post","status-publish","format-standard","hentry","category-qc-sop"],"_links":{"self":[{"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/posts\/812","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=%2Fwp%2Fv2%2Fcomments&post=812"}],"version-history":[{"count":1,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/posts\/812\/revisions"}],"predecessor-version":[{"id":813,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/posts\/812\/revisions\/813"}],"wp:attachment":[{"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=%2Fwp%2Fv2%2Fmedia&parent=812"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=%2Fwp%2Fv2%2Fcategories&post=812"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=%2Fwp%2Fv2%2Ftags&post=812"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}