{"id":797,"date":"2025-06-14T14:38:10","date_gmt":"2025-06-14T14:38:10","guid":{"rendered":"https:\/\/pharmacoguide.com\/?p=797"},"modified":"2025-06-14T14:38:10","modified_gmt":"2025-06-14T14:38:10","slug":"sop-title-operation-cleaning-and-calibration-of-muffle-furnace","status":"publish","type":"post","link":"https:\/\/pharmacoguide.com\/?p=797","title":{"rendered":"SOP Title: Operation, Cleaning and Calibration of Muffle Furnace"},"content":{"rendered":"<p><strong><u>SOP Title:<\/u><\/strong> <strong><u>Operation, Cleaning and Calibration of Muffle Furnace<\/u><\/strong><\/p>\n<ol>\n<li><strong>Objective:<\/strong><\/li>\n<\/ol>\n<p>To lay down the procedure for Operation, Cleaning and Calibration of Muffle Furnace.<\/p>\n<ol start=\"2\">\n<li><strong>Scope:<\/strong><\/li>\n<\/ol>\n<p>This SOP is applicable for Operation, Cleaning and Calibration of Muffle Furnace located in Quality Control Laboratory<\/p>\n<p><strong>\u00a0<\/strong><\/p>\n<ol start=\"3\">\n<li><strong>Responsibility:<\/strong><\/li>\n<\/ol>\n<ul>\n<li>Quality Control: To prepare and review the SOP. To follow the procedures laid down for Operation, Cleaning and Calibration of Muffle Furnace as per this SOP.<\/li>\n<li>Engineering Department: To carry out preventive maintenance as per schedule and procedure defined.<\/li>\n<li>Quality Assurance Department: To review and approve the SOP and Annexures.<\/li>\n<\/ul>\n<p><strong>\u00a0<\/strong><\/p>\n<ol start=\"4\">\n<li><strong>Accountability:<\/strong><\/li>\n<\/ol>\n<p>Head Quality Control Department, Head Quality Assurance Department.<\/p>\n<p>&nbsp;<\/p>\n<ol start=\"5\">\n<li><strong>Procedure:<\/strong><\/li>\n<\/ol>\n<p><strong>\u00a0<\/strong><\/p>\n<table width=\"691\">\n<tbody>\n<tr>\n<td width=\"67\"><strong>\u00a0\u00a0\u00a0 5.1<\/strong><\/td>\n<td colspan=\"2\" width=\"624\"><strong>Safety Precautions\/Instructions:<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.1.1<\/td>\n<td width=\"607\">Ensure all the electrical connections are properly earthed. Ensure the power plug of the muffle furnace is fixed in the socket properly.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.1.2<\/td>\n<td width=\"607\">Wear personal protective equipments such as safety glasses, heat resistant gloves during operation of equipment.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.1.3<\/td>\n<td width=\"607\">Ensure that the ON\/OFF switch is working properly by switching \u201cON\u201d the muffle furnace.\u00a0 <strong>\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0\u00a0<\/strong><\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">\u00a0\u00a0\u00a0 <strong>5.2<\/strong><\/td>\n<td colspan=\"2\" width=\"624\"><strong>Operation:<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.1<\/td>\n<td width=\"607\">Switch ON the main switch.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.2<\/td>\n<td width=\"607\">The main indicator red light and yellow light will glow and the temperature panel display will show the temperature in <sup>o <\/sup>C.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.3<\/td>\n<td width=\"607\">Set the required temperature on the digital temperature controller by pushing red key present at the left side of \u201cTemp. SET\u201d knob and rotating the \u201cTemp. SET\u201d knob.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.4<\/td>\n<td width=\"607\">Relieve the red key button after the required temperature is set.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.5<\/td>\n<td width=\"607\">Move the \u201cEnergy Regulator\u201d knob clockwise to corresponding temperature.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.6<\/td>\n<td width=\"607\">Wait till the temperature controller display shows the set temperature.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.7<\/td>\n<td width=\"607\">Open the door of the muffle furnace by pushing the lever upwards and place the crucible containing sample inside the muffle furnace with the help of long pair of tongs.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.8<\/td>\n<td width=\"607\">Immediately close the door.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.9<\/td>\n<td width=\"607\">Allow the substance under test to remain inside muffle furnace at a specified temperature and specified time as mentioned in the respective pharmacopoeia or specification.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.10<\/td>\n<td width=\"607\">After completion of analysis, Switch OFF the mains of the muffle furnace.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.11<\/td>\n<td width=\"607\">Open the door of the muffle furnace, remove the crucible containing sample and place it in desiccator till it attains the room temperature.<\/td>\n<td width=\"17\"><strong>\u00a0<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.12<\/td>\n<td width=\"607\">Immediately close the door of the muffle furnace.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.2.13<\/td>\n<td width=\"607\">Update the \u201cInstrument\/Equipment Usage Log Book\u201d as per SOP \u201cCalibration of Instruments and Equipment\u2019s\u201d.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\"><strong>5.3<\/strong><\/td>\n<td colspan=\"2\" width=\"624\"><strong>Cleaning Procedure:<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.3.1<\/td>\n<td width=\"607\">Switch OFF the mains of the muffle furnace.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.3.2<\/td>\n<td width=\"607\">Ensure that the temperature of the muffle furnace is nearly at room temperature when cleaning inside.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.3.3<\/td>\n<td width=\"607\">Clean the inside and outside of muffle furnace with lint free cloth or soft brush before and after use.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\"><strong>5.4<\/strong><\/td>\n<td colspan=\"2\" width=\"624\"><strong>Calibration Procedure:<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"67\">&nbsp;<\/td>\n<td colspan=\"2\" width=\"624\"><strong>Frequency: <\/strong><strong>\u00a0<\/strong>Once in a year by external agency.<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.4.1<\/td>\n<td width=\"607\">Calibration is carried out by external agency as per SOP \u201cCalibration of Instruments and Equipment\u2019s\u201d.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.4.2<\/td>\n<td width=\"607\">Refer Annexure for calibration test to be performed and acceptance criteria.<\/p>\n<table width=\"82%\">\n<tbody>\n<tr>\n<td width=\"12%\"><strong>Sr. No.<\/strong><\/td>\n<td width=\"25%\"><strong>Tests<\/strong><\/td>\n<td width=\"28%\"><strong>Set value<\/strong><\/td>\n<td width=\"33%\"><strong>Acceptance Criteria <\/strong><\/p>\n<p><strong>(\u00b1 2%)<\/strong><\/td>\n<\/tr>\n<tr>\n<td rowspan=\"5\" width=\"12%\">1.<\/td>\n<td rowspan=\"5\" width=\"25%\">Temperature Verification<\/td>\n<td width=\"28%\">200<sup>o<\/sup>C<\/td>\n<td width=\"33%\">\u00b1 4<sup>o<\/sup>C<\/td>\n<\/tr>\n<tr>\n<td width=\"28%\">400\u00b0C<\/td>\n<td width=\"33%\">\u00b1 8<sup>o<\/sup>C<\/td>\n<\/tr>\n<tr>\n<td width=\"28%\">800<sup>o<\/sup>C<\/td>\n<td width=\"33%\">\u00b1 16<sup>o<\/sup>C<\/td>\n<\/tr>\n<tr>\n<td width=\"28%\">1000<sup>o<\/sup>C<\/td>\n<td width=\"33%\">\u00b1 20<sup>o<\/sup>C<\/td>\n<\/tr>\n<tr>\n<td width=\"28%\">1100<sup>o<\/sup>C<\/td>\n<td width=\"33%\">\u00b1 22<sup>o<\/sup>C<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.4.3<\/td>\n<td width=\"607\">Generate the A. R. No. on the \u201cInstrument\/Equipment Calibration Register\u201d as per SOP and give requisition to QA department for calibration formats of instruments\/equipments on the due date of calibration.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.4.4<\/td>\n<td width=\"607\">For issuance of calibration report refer SOP \u201cControl of documents, Preparation, Approval, Issuance and Maintenance.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.4.5<\/td>\n<td width=\"607\">Update \u201cAnnual Calibration Schedule\u201d and \u201cMonthly Calibration Schedule\u201d as per SOP \u201cCalibration of Instruments and Equipments\u201d.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\"><strong>5.5<\/strong><\/td>\n<td colspan=\"2\" width=\"624\"><strong>Preventive Maintenance:<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.5.1<\/td>\n<td width=\"607\">Perform the preventive maintenance as per define Annexure.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.5.2<\/td>\n<td width=\"607\">If preventive maintenance activity is outsourced, the observations are to be transcribed in the Annexure and external party certificate to be attached.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<tr>\n<td width=\"67\">5.5.3<\/td>\n<td width=\"607\">If any breakdown occurs, follow procedure as per SOP \u201cCalibration of Instruments and Equipments\u201d.<\/td>\n<td width=\"17\">&nbsp;<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p><strong>\u00a0<\/strong><\/p>\n<ol start=\"6\">\n<li><strong>Definitions<\/strong><strong> \/ Abbreviations:<\/strong><\/li>\n<\/ol>\n<ul>\n<li><strong>Definitions:<\/strong><\/li>\n<li>Calibration: Tests performed to ensure that the instrument\/equipment is functioning as per set parameters. These tests are performed as per predefined frequency and also after any major repair or replacement of parts.<\/li>\n<li>Preventive Maintenance: Maintenance activity performed to ensure that the equipment will function smoothly and to avoid breakdowns. These activities are performed as per predefined frequency.<\/li>\n<li><strong>Abbreviations<\/strong><\/li>\n<\/ul>\n<p><strong>\u00a0<\/strong><\/p>\n<table>\n<tbody>\n<tr>\n<td width=\"116\"><strong>Abbreviation<\/strong><\/td>\n<td width=\"276\"><strong>Expansion<\/strong><\/td>\n<\/tr>\n<tr>\n<td width=\"116\">SOP<\/td>\n<td width=\"276\">Standard Operating Procedure<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">QC<\/td>\n<td width=\"276\">Quality Control<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">QA<\/td>\n<td width=\"276\">Quality Assurance<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">A. R. No<\/td>\n<td width=\"276\">Analytical Reference Number<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">No.<\/td>\n<td width=\"276\">Number<\/td>\n<\/tr>\n<tr>\n<td width=\"116\"><sup>o<\/sup>C<\/td>\n<td width=\"276\">Degree Celsius<\/td>\n<\/tr>\n<tr>\n<td width=\"116\">PM<\/td>\n<td width=\"276\">Preventive Maintenance<\/td>\n<\/tr>\n<\/tbody>\n<\/table>\n<p>&nbsp;<\/p>\n<p><strong><em><u>*Note &#8211; Ready to use SOP available in \u201cDOWNLOAD\u201d Section.<\/u><\/em><\/strong><\/p>\n<p>&nbsp;<\/p>\n","protected":false},"excerpt":{"rendered":"<p>SOP Title: Operation, Cleaning and Calibration of Muffle Furnace Objective: To lay down the procedure for Operation, Cleaning and Calibration of Muffle Furnace. Scope: This SOP is applicable for Operation, Cleaning and Calibration of Muffle Furnace located in Quality Control Laboratory \u00a0 Responsibility: Quality Control: To prepare and review the SOP. To follow the procedures [&hellip;]<\/p>\n","protected":false},"author":1,"featured_media":0,"comment_status":"open","ping_status":"closed","sticky":false,"template":"templates\/template-full-width.php","format":"standard","meta":{"footnotes":""},"categories":[4],"tags":[],"class_list":["post-797","post","type-post","status-publish","format-standard","hentry","category-qc-sop"],"_links":{"self":[{"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/posts\/797","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/posts"}],"about":[{"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/types\/post"}],"author":[{"embeddable":true,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/users\/1"}],"replies":[{"embeddable":true,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=%2Fwp%2Fv2%2Fcomments&post=797"}],"version-history":[{"count":1,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/posts\/797\/revisions"}],"predecessor-version":[{"id":798,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=\/wp\/v2\/posts\/797\/revisions\/798"}],"wp:attachment":[{"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=%2Fwp%2Fv2%2Fmedia&parent=797"}],"wp:term":[{"taxonomy":"category","embeddable":true,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=%2Fwp%2Fv2%2Fcategories&post=797"},{"taxonomy":"post_tag","embeddable":true,"href":"https:\/\/pharmacoguide.com\/index.php?rest_route=%2Fwp%2Fv2%2Ftags&post=797"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}